berlinger SmartMonitor SITE L Quick Info
Hide thumbs Also See for SmartMonitor SITE L:

Advertisement

Quick Links

smartsystem

SmartMonitor information

i

Device layout

11
10
9
8
State of delivery
You receive the SmartMonitor in sleep mode.
7
The display is turned off.
1 x short press to wake up the device.
Centered hook for mounting on device backside.

Device installation

1
With external sensor
1. Position the external sen-
sor inside (1) and the Smart-
Monitor device outside (2)
the refrigerator/freezer and
wait for the external sensor
to acclimatize.
Attention: The SmartMonitor SITE L requires
permanent charging with the delivered power
How to configure your
2
device

Automatic configuration

1. Short-press the multifunctional button to
"wake up" the SmartMonitor from sleep mode.
Do not apply force as it will damage
the multifunctional button.

Manual configuration

If the SmartMonitor is not able to establish
connectivity with SmartView to download
the configuration, it is required to manually
download the configuration file to the device
via USB.
Follow the instructions in the user manual to
set up the configuration manually.
Need help with Manual configuration?
See user manual
«How to configure your
1
2
3
4
5
6
2. Plug in the USB-cable
into the SmartMonitor and
the power adapter into the
power socket for permanent
charging.
adapter.
2. To configure the device
with the desired config-
uration, triple push the
multifunctional button in
quick succession.
This will initiate cloud syn-
chronization to download
the configuration.
device».
QUICK INFO
SmartMonitor SITE L
1
Display
2
USB-plug
3
LED
4
Removeable cap
5
Device use
6
Device type
12
Available icons on device display
Battery charging
Battery status
Cloud status
Warning
symbol
3. Remove the grey cap
on the right side of the
SmartMonitor device.
Warning: Do not apply
force!
5. Once correctly applied,
tighten the connector by
screwing clockwise.
The screen is indicating that the configuration
download is initiated and all display icons are
loading. This might take several seconds.
If the display indicates MANUAL, automatic
For more information scan the QR code or visit
www.berlinger.com/user-manual-smartmonitor.
7
Serial number
GTIN barcode
8
Multifunctional button
9
10
Light sensor
11
Humidity sensor
12
Sensor socket (M5)
Cellular connection
Alarm status
WiFi
USB
Bluetooth
4. Align the sensor plug
with the sensor socket.
Ensure the Berlinger
eye is facing the
same direction as the
SmartMonitor display.
Now, plug the external
sensor into the sensor
socket.
While the device is
establishing connection
to SmartView, the tick in
the cloud symbol ( ) is
blinking and the cellular
symbol ( ) is static.
Once the download is
successfully completed,
the configuration policy
name of your desired
confiugration policy is
displayed.
setup was not successful.
Learn More
User Manual SmartMonitor
Measurement
status
7-digit text field

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SmartMonitor SITE L and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for berlinger SmartMonitor SITE L

  • Page 1 5. Once correctly applied, tighten the connector by screwing clockwise. Attention: The SmartMonitor SITE L requires permanent charging with the delivered power adapter. How to configure your The screen is indicating that the configuration download is initiated and all display icons are device loading.
  • Page 2 See user manual «How to start your device». Running device Stop monitoring SmartMonitor SITE L can be stopped via cloud Example of an OK display command only. Device indicates that To stop monitoring, log into SmartView. there is no temperature...

Table of Contents