Page 1
Dual-Column Cloud Condensation Nuclei Counter Operator Manual for Dual-Column CCN DOC-0128 Revision F 2400 Trade Centre Avenue Longmont, CO 80503 USA...
Page 2
Droplet Measurement Technologies. Although every precaution has been taken in the preparation of this document, Droplet Measurement Technologies assumes no responsibility for errors or omissions. Neither is any liability assumed for damages resulting from the use of the information contained herein.
C O N T E N T S Setting up the Dual-Column CCN (CCN-200) ............8 Unpacking ..........................8 Connecting the Instrument Components ................9 USB Memory Stick ....................... 11 Setting the Flow and Humidifying the CCN-200 ..............12 Checking out the CCN-200 ....................14 1.5.1...
Setting up the Dual-Column CCN (CCN-200) Several steps are required for initial CCN-200 set-up: • Unpacking and mounting the instrument Connecting CCN components such as the power cables, supply and drain bottles, and touch • screen monitor • Setting the flow and humidifying the instrument (humidification can take up to 12 hours and must be performed every time the CCN is dry when started) •...
Page 17
If additional water has entered the system, contact Droplet Measurement Technologies immediately. WARNING! The CCN Counter must be prepared for shipping by drying the column and draining the OPC.
The CCN-200 (Figure 9) is based on the design of Dr. Greg Roberts of Scripps Institute of Oceanography. Dr. Athanasios Nenes of Georgia Institute of Technology did the CCN-200’s theoretical parameterization. The technology is patented, and Droplet Measurement Technologies possesses the exclusive license to manufacture the instrument.
Activated droplet counting is done with two Optical Particle Counters (OPCs), one for each column. The OPC has been specially designed for the CCN and uses side-scattering technology for the particle sizing.
Operator Manual, Dual-Column Cloud Condensation Nuclei Counter (CCN-200) CCN Maintenance Droplet Measurement recommends the following service and maintenance schedule for the Dual Column CCN. The maintenance schedule is based on the continuous operation of the instrument and may need to be modified for other operating conditions.
5. Note the bin where the calibration particles are measured. A good calibration exists if the particles are in the proper bin or slightly high. If the particles are in a lower size bin, the OPC could be dirty and will need cleaning and recalibration. Contact Droplet Measurement Technologies.
Page 101
Down should only be selected when the unit needs to be totally dried out prior to shipping. Inlet Set (C): The set point temperature at the CCN-200 inlet manifold. Köhler Curve: A mathematical function representing the cloud-droplet formation process. Köhler curves are used to calibrate the CCN-200 and determine the relationship between temperature gradient and supersaturation.
DMT will have a “tested” label on them. Please contact Duncan Axisa at Droplet Measurement Technologies if you have any questions or concerns, or if you need to order replacement filters. Email CCN-Info@dropletmeasurement.com; Phone: 303-440- 5576 EX.
Need help?
Do you have a question about the Dual-Column CCN and is the answer not in the manual?
Questions and answers