MiniSonic _ P
MiniSonic _ P-B
Portable Flowmeters
1 / 26

Advertisement

Table of Contents
loading

Summary of Contents for Ultraflux MiniSonic P

  • Page 1 Tel: +33 (0) 1 39 79 26 40 Fax : +33 (0) 1 39 79 91 22 E-mail : ultraflux@ultraflux.fr Web : www.ultraflux.com MiniSonic _ P Ultraflux NT 214B GB1 MiniSonic _ P-B Portable Flowmeters Revision : 26/08/2008 NT214B GB1...
  • Page 2 PREAMBLE : Thank you for choosing Ultraflux to make your flow measurement. We offer a full range of portable or fixed flowmeters backed by 30 years’ expertise and experience using Ultrasonic techniques: ® The Digisonic and UF 801-P families of Portable flowmeters with integral loggers and signal analysis functions ®...
  • Page 3: Table Of Contents

    SUMMARY : 1 – Typical Applications: Liquid flow measurements and controls Clamp on or insertion transducers Use for official control or as a standard for calibration 2 – Composition of a measurement point. Certifications Applicable standards MiniSonic limitations Recommended accessories 3 –...
  • Page 4: Typical Applications

    1 – TYPICAL APPLICATIONS MiniSonic P and P-B main purpose is to offer a performing flow measurement solution at a very competitive price by saving any installation costs. With clamp-on transducers, there is no need of any spool piece or tapping by keeping the same final accuracy than any flowmeter.
  • Page 5 1-2 - The clamp on solution is available for most of the liquids flowing in metallic or plastic pipes. Ultraflux offers a very wide transducers range with different sizes and frequencies to be selected depending on the application and pipe diameter from 10 mm I.D. to more than 3 metres.
  • Page 6: Composition Of A Measurement Point

    2 - COMPOSITION OF A MEASUREMENT POINT - The converter MINI-P or P-B with its carrying bag. - Its standard accessories: batteries (NIMH) charger and cable - One or more set of two probes depending on your request / applications - Integral or optional supports (SU - Coupling gel (BT = low temperature / HT = high temp.) for clamp-on probes - Probe / converter link cables (L = 2.5 m) and optionally extensions (L= 5 m)
  • Page 7: Ergonomics And Dimensions

    3 - ERGONOMICS AND DIMENSIONS OF THE MINISONIC-P and P-B CONVERTERS All operations can be done while keeping MiniSonic–P inside its protection bag. ® MINISONIC P interfaces location view Pipe 2 Connectors to Probes Mini P-B Cable with a Y ending...
  • Page 8 4 –USUAL CONNECTIONS ON MINISONIC-P or P-B : When there is no need of automatic records, connections as described in § 3 are sufficient. > To permit records and connection to an external data-logger or to a PC , or to any recorder or transmission device , please use our cable and add if necessary connections on 15 pins male connector .
  • Page 9 > Automatic records by using our Software on a desk top or lap top PC : Please refer to our LS 600W x x Software manual (CD Rom) - Once this software installed on your laptop PC you can set in its configuration menu a sampling period and also the duration of the records.
  • Page 10: Converter First Handling Instructions

    5 – CONVERTER FIRST HANDLING INSTRUCTIONS: ® Charging Batteries - Charger/ Adapter + Auxiliary supply Cords : EU – Optionally other ( US …) Approval = CE and CB/UL AC Input Voltage : 100 -240 V 47/63Hz Output Voltage : 15 V DC - 1 A Charging time = 14 hours A green LED is On when supplied.
  • Page 11: Probes Installation

    – General Procedures. The MiniSonic P & P-B accept all standard or special transducers of Ultraflux catalogue with frequency from 250 KHz up to 2 MHz : clamp-on , insertion or wetted … ® Unless specified and instructed otherwise in the order, refer to the probe data sheet of the model used .
  • Page 12 Probes installation : particular recommendations for a portable use : Please read general recommendations as detailed in our training manual NT 122. Under header § 2, we give some information about expected accuracy depending on available straight length and transducers mounting modes. When you will arrive in front a new pipe to be measured, you must have some good reflex, and your experience will help for.
  • Page 13: Measurement Display Menu

    1. Once the cabling has been carried out and verified, power up the system: > The MiniSonic must display its home screens: - Hardware version = MiniSonic P or P-B and P.C. Board version (MK4_M). - Firmware version = 18.10-01 (e.g. – make a note of the first figure)
  • Page 14 DETAILED REVIEW OF MENUS – MiniSonic Versions P and P-B : Here below we list first the standard P version and at the end of this chapter 8 , in 8-11, we had the extra screen or readings concerning the P-B version . 8 -1 : “...
  • Page 15: Calibration Allowed Menu

    It will be required for any subsequent intervention. ® In case of loss, contact Ultraflux giving the following information’s; the serial number of your MiniSonic MK4*/xx/xx/xxxx as it appears during energizing and the date or dates on which you want action to take place. An operation will be requested, allowing a provisional code to be calculated.
  • Page 16 8-4 : « Q Settings » menu ( Flow / Fluid ) ® Following data are common to the two pipes. Q unit = m ∇ Choice of Flow unit: ( l/h …tot … m3/s ); ∇ Qmax (unit) Scale (graphics on PC only): +/xxxxx.x Outputs: see menu «...
  • Page 17 8-5 : « Probe / Echo parameter settings » menu : ∇ Probe = SE xxxx Choice of probe (or SM ou SP. A/B) From this choice, it results Emitting Frequency, D.S. calculation (S.E.) or Opening of L and D fields (SM) ∇...
  • Page 18 8-6 : « General Settings » menu ∇ Back Light Setting of LCD back-lighting ON / OFF / TIMED (Depending on MiniSonic version ) ∇ Filter = ∗∗ Filter rate of measurements (Number of 0.5 s cycles) ∇ Mémory (s)= ∗∗ Last accepted measure (IQ >...
  • Page 19: Factory Settings Menu

    ! Caution = to carry out Self-zeroing, the flow rate must be absolutely zero while the display indicates « * dTo * » . 8-8 : « Output Current Settings» menu ( normally reserved to Ultraflux ) It permits adjustment of coefficients 4 mA ( 1 &...
  • Page 20 8-11 : Optional menus or readings concerning MiniSonic P-B two-pipes version The optional menus concern the second pipe and permit to validate the second probes connector. The first and more important is in the Echo menu : Nb of Chords = 1 or 2 The MiniSonic P-B can calculate a Q result from Q1 &...
  • Page 21 Each value or text field is filled in by Ultraflux before the final tests and delivery. Nevertheless, it is advisable to verify them by running through the various menus, comparing them with files given at the time of order, or with your instructions or particular choices.
  • Page 22 • If the MiniSonic displays INIT constantly, attempt a power cutoff, then re-energize with if possible charger connected. If the fault persists, contact Ultraflux (Problem with the microprocessor). • If the MiniSonic displays “Flow Rate Fault”, this message does not mean that the MiniSonic has failed, but that the ultrasound signals have not reached the expected level or are outside the window [Co +/- Delta C].
  • Page 23 • Or measuring instruments for dimensional or alignment measurements SPARE PARTS : Ultraflux offers possibilities of purchasing spare cards or, as long as acceptance is granted, the standard replacement of equipment in the event of failure. Contact the Ultraflux commercial services or the regional dealer.
  • Page 24: Appendix

    11 – APPENDIX 1 : Calibration file example with a MiniSonic-P-B with clamp-on probes SE NT214B GB1 24 / 26...
  • Page 25 APPENDIX 2: Speed of sound in fluids – Relation with density + P & T Influence The speed of sound in a fluid is a full-fledged physical characteristic which may be linked with others, such as density (d), pressure (P), temperature (T) or the % of two or several miscible products or with another magnitude having good analogy, like compressibility.
  • Page 26 Superheated water involves curve or graph networks taking into consideration the pressure and the temperature: consult us. Measurement on petroleum products – values at 15°C / 1 bar (except for butane and propane): 1 5 0 0 W a te r 1 4 0 0 D o m e s tic fu e l o il G a s O il...