Menu item Option
Device
pHix pH
type
Electrode
pH
type
Zero
pH-
point 4.6
Electrode
pH
Zero
zero type
point 7.0
pH
Serial
number
Part
number
Sensor on-
chip serial
number
Number &
Revision
Compile
time
Slope
Zero point
pHix Redox
pHix Configurator software /Main configuration
 
 
Description
The sensor type
The pH electrode type
The sensor serial number
The sensor item number
Chip serial number
The current firmware in the sensor
The last time the firmware was changed
The slope is determined by measuring the difference in mV
readings pH buffers. The ideal slope should be approximately
-59.16 mV/pH at 25°C
The pH where the Zero Point is 0 mV, it means the same
amount of H+ and OH- ions.