Electrode Storage; Specifications - Xylem MJK pHix Compact V3 NG Manual

Table of Contents

Advertisement

Note! For further accessories, see pHix Compact Datasheet which can be downloaded
from our
Download Center

Electrode storage

When a pH-electrode is not in use and is kept in storage, always put the electrode in a
3mol KCL-solution. If a storage solution is not available, a buffer 4 can be used
temporarily.
Note! Never store the electrode in demineralized water.

Specifications

pHix® Compact
Supply voltage
Consumption
Temperature range
Housing material
Enclosure
Input
Input, impedance
Output
Temperature comp.
Temp. electrode
Buffer adjustment
Cable length
https://mjk.com/download-center?mainCatId=3104&catId=all&lang=english
2
Maintenance /Electrode storage
 
 
2
under Datasheet.
10-28 V DC
App. 50mW
-4 ... +176° F / -20 ... +80° C 
Housing exists in two variants:
1.
with PPS Fortron only
2.
with PPS Fortron and stainless steel (EN
1.4404 / ASTM 316L) combined.
NEMA 6X / IP 68, withstands submersion to 10
mWG
0-14 pH/-1000 ... +1000 mV
Measuring input: > 10
Reference input: > 10
0-14 pH / 4-20 mA.
32-122° F / 0-50° C = 4-20mA
 Built-in sensor. Entire operating range  
Built-in to electrode holder
Automatic with tilt switch or switch ring
 19.7 feet / 6 meter
pHix® Compact V3. NG (next generation)
12
ohm.
6
ohm
EN 4.10 Version 2202
31

Advertisement

Table of Contents
loading

Table of Contents