Temperature Sensor Troubleshooting

If the Temperature sensor is reading 255°F, the Temperature sensor
may be disconnected from the Control Module, may not be initialized, or
may be defective.
Check the cable from the sensor to the Control Module for any cuts or
kinks. Check the cable connection to the Control Module.
The initialization procedure should be performed when replacing or
adding a sensor or when a sensor has lost its initialization to the Control
Module and is reading 255°F.
To initialize a temperature sensor to the Control Module, perform the
following procedure:
1. Disconnect power from the Control Module
2. Connect the Temperature Sensor to the appropriate
Temperature Channel on the Control Module
3. Reapply power to the Control Module
4. Verify that the Temperature reading is correct
Columbia Weather Systems, Inc.
89