Page 1
Handle this product in accordance with the explanations described in this manual. After reading this operation manual, keep it with the warranty in a safe place. If you find any unclear points in this manual, please contact your Sibata distributor. SIBATA SCIENTIFIC TECHNOLOGY LTD.
△ [Modifications and Repairs Are Prohibited] Never disassemble or modify this product, as doing so will void the warranty with Sibata. Doing so may also result in unforeseen faults and accidents. △ If a fault occurs, request repairs immediately. Never leave the fault as is, or attempt your own repairs, as subsequent use may result in unforeseen accidents.
1 Overview This Andersen Sampler equipped with multi-stage, multi-orifice nozzles is used to measure particle size distribution of airborne aerosol in air and the environment by the impactor method. It is configured with a classifier main unit, flowmeter, pump, and shelter. 1-1 Principle The Andersen Sampler has a structure where usually eight corrosion-resistant aluminum alloy stages are stacked as shown in Fig.
Page 5
Formula (3) is substituted into formula (2) and each stage's 50 % separated dynamic particle diameter d is obtained by the following formula. μ π Ψ × ρ − × × ··············· (4) However, Q = 28.317 cm /min = 28.3 L/min ρ...
2-2 Andersen Sampler AN-200 Set < Item code 080040-01 > Shelter Andersen Sampler AN-200 main unit Flowmeter Power switch Suction pump Fig. 3 2-3 Andersen Sampler AN-200 Main Unit < Item code 080040-012 > Flowrate adjustment valve Andersen main unit Float type area flowmeter Fig.
3 Operational Procedures 3-1 Preparation (1) Thoroughly wash collection plates and dry them using a desiccator or something similar. (2) Measure the weight of the collection plates described in (1) and the backup filter. 3-2 Operation (1) Place the backup filter on the backup filter holder of the Andersen Sampler. (2) Place a collection plate on each stage of the sampler.
4 Sample Evaluation Method 4-1 Mass Measurement The mass is calculated from the difference between the mass of the collection plates before sampling and after sampling. For more accurate measurement, a precision balance (minimum scale line of 0.01 mg or less) needs to be used. Measurement in a thermostatic chamber is more desirable when a precision balance is used.
5 Maintenance and Inspections 5-1 Cautions for Use (1) Prevent any hose from being bent along its length or clogged. (2) Secure sufficient ventilation during continuous operation. (3) Dry run the unit (approx. 30 min) every two weeks when you do not plan to use it for a long time.
Page 11
(3) Remove the four knurled nuts fixing the suction pump. (4) Remove the ground wire (green). (5) Remove the suction pump from the unit. At this point, remove the spacers attached to the threads as well. Be careful not to lose the spacers. Spacer (6) Perform the reverse procedures of the procedures above to install a new pump in the unit.
6 Specifications Model AN-200 Particle Classification 0.43 to 11 μm Range Features Particle Size Classification 8 levels (stage) ・ Study of air pollution ・ Classification of radioactive aerosol Application ・ Study of respiratory disorders ・ Evaluation of air filter efficiency Suction Flowrate 28.3 L/min Maximum Flowrate...
7 Warranty If a Sibata product fails within 1 year from date of purchase, it will be repaired free of charge. To request repairs, contact your Sibata distributor. Be sure to provide the item code, product name, model number, serial number, a description of the problem, and other relevant information.
Page 14
Note) Shape, dimensions, specifications, and other product information are subject to change without notice in the interest of product improvement to the extent that product functions and applications will not be impaired.