CODE 090130-31 High Volume Air Sampler HV-500R series OPERATION MANUAL Thank you for purchasing this product. This operation manual describes precautions that are important for preventing accidents as well as the procedures used to handle the product. To ensure safety, read this operation manual and the attached warranty thoroughly before use, and use the product correctly.
Contents Before Use ····································································································· 1 ■ About this operation manual ······································································· 1 ■ Checking the package ················································································ 1 Safety Precautions ·························································································· 2 ■ About the User (Important) ··········································································· 2 ■ Warning Labels ························································································· 2 Product Overview ···························································································· 5 ■ Features ··································································································· 5 ■...
If you discover any errors or omissions, however, please contact your Sibata representative. The copyright of this operation manual belongs to Sibata Scientific Technology Ltd. The reproduction of all or part of this operation manual without prior written permission from Sibata Scientific Technology Ltd. is prohibited. Checking the Package Check the contents of the package before using the product.
Safety Precautions The precautionary information that appears in this operation manual is for ensuring that the product is used safely and for preventing injury to you and other people and damage to equipment. It is all important for ensuring safety and so be sure to read it thoroughly before using the product and observe it during use.
Page 5
When the abnormality is judged to be caused by this product, turn the power switch OFF, disconnect the power plug and contact your Sibata agent. Do not use this product in an abnormal state or allow it to be dismantled for repair by non-service personnel. Doing so might cause malfunction or accidents.
Page 6
Doing so might damage the hardware, for example. Note that, should some nonconformity occur, Sibata does not assume any liability whatsoever for compensation of data or content that could not be acquired or logged as a result, loss of data or other content, and other direct and indirect damages relating to the preceding.
Product Overview The HV-500R is sample suspended dust in the open face, and provide quantitative analysis of dust concentration and composition. High-accuracy flow rate control is provided by a differential pressure detection system, and instantaneous and cumulative flow rate values are displayed digitally. Reductions in suction flow rate due to increased dust collection are minimized thanks to a constant flow rate function.
A special orifice flow meter can be attached to perform flow rate calibrations. Others We accept special orders for additional connecting parts and special measurements. Contact your Sibata agent. Product Limitations This product is a suction based sampler. Note that it is not capable of the following: ...
Name of Parts Front・Suction side ⑨ ⑧ ③ ② ④ ① ⑤ ⑥ ⑦ ⑪ ⑩ ① Suction port ② Atmospheric pressure sensor port ③ Outside air temperature sensor ④ Clip holder ⑤ Filter clip (standard accessories) (The above figure has a filter.) ⑥...
Operation Preparation Replacement of the collecting filter Set a φ110mm Filter no a Filter clip. Front side of the filter turns up as illustrated below. Clip ring Convex (Two points) φ110mm Filter Front up Filter clip Slit (2 places) ※...
Wiring Methods Use a single phase 220 V power supply. (180 V to 264 V can be used.) Be sure to ground this product. If you wish to use a non-specified voltage, contact your Sibata agent. Green : Grounding wire...
Touch panel protection sheet A Touch panel protection sheet is enclosed. Please put the protection sheet on the touch panel if any damages on the touch panel are concern depending on a sampling situation. Main Screen When the power is first turned ON, "System Initializing" will be displayed for 5 seconds. The screen then changes to the following screen: High Volume Air Sampler Model name...
Sampling Method Setting the Flow Rate Touch the numerical region of the flow rate 0 9 / 0 3 1 3 : 4 0 : 0 0 25deg settings directly to proceed to the flow rate setting FLOW SET 500 L/min screen, where the flow rate can be set.
Page 15
Setting the Operation Start Timer Press the characters in the operation start timer area to switch between the delay timer and the clock timer. The delay timer, shown onscreen as [DELAY], starts the operation after a certain period of time has elapsed. The clock timer, shown onscreen as [CLOCK], starts the operation at a specific hour and minute.
Page 16
25deg Press the numbers next to [CLOCK] to proceed to 0 9 / 0 3 1 3 : 4 0 : 0 0 the start time setting screen, where the starting time FLOW SET 500 L/min can be set. ST. CLOCK 09/03 14:00 * If the numbers next to [CLOCK] show a time SAMP.
Page 17
Setting the Operation End Timer Press the characters in the operation end timer area to switch between the sampling timer and the volume timer. The sampling timer, shown onscreen as [TIMER], ends the operation after the configured period of time has elapsed. The volume timer, shown onscreen as [VOL.], ends the operation when the configured cumulative flow rate has been reached.
Page 18
Press the numbers next to [VOL.] to proceed to the 25deg 0 9 / 0 3 1 3 : 4 0 : 0 0 ending cumulative flow rate setting screen, where the ending cumulative flow rate can be set. FLOW SET 500 L/min START DELAY 01hr00min * Entering 0.0 m...
Page 19
Starting Operations In the main screen, confirm the setting conditions. In the example in the figure below, the set flow rate is 1,000 L/min, operations start 1 hour later, and operations continue for 8 hours. 25deg 0 9 / 0 3 1 3 : 4 0 : 0 0 FLOW SET 500 L/min...
Page 20
After the [START/STOP] button has been pressed, if the flow meter inadvertently responds to a slight airflow, a zero error will occur and the following screen will be displayed. ZERO ERROR 23.2 Press [ESC] to return to the main screen. Then press the [START/STOP] button again. * A zero error determination will be made when the [START/STOP] button is pressed, even under timer operation.
Operation completed When an operation is finished, an Integrated Flow Volume is displayed as shown below. 0 9 / 0 3 1 3 : 4 0 : 0 0 25deg TOTAL MODE ESC Key to back to Main display. Error Displays If an error occurs during operation, this product will log the error.
Power Outages If a power outage occurs during operation, the product will turn OFF and stop operating. However, the remaining sampling will continue after power is restored. E.g.) If an 8 hour sampling period has been set, and a 1 hour power outage occurring during this time, sampling will be delayed by 1 hour, but the full 8 hours of sampling will be completed.
When the internal battery run-down icon is displayed, the battery needs to be changed immediately. Please contact your agent or Sibata Scientific Technology Ltd. Mode Display In the main screen and other screens, press [MODE] to view environmental information. The displayed items will differ before and after blower operation.
Page 24
During Standby (after the [START/STOP] button is pressed but before the blower runs) Press [MODE] to toggle in sequence between the screens shown below. Press [ESC] to return to the main screen. 0 9 / 0 3 1 3 : 4 0 : 0 0 25deg 0 9 / 0 3 1 3 : 4 0 : 0 0 25deg...
Menu In the main screen, press [MENU] to display the menu shown in the following figure. Sampling data SAMP. DATA STD SET Toggling flow rate correction method 25deg 1atm Time setting RTC SET Error setting DISPLAY TRBL SET Screen setting SERIAL No.
Page 26
Toggling the Flow Rate Correction Method Press this key to toggle between 20 °C 1 atmospheric correction, 25 °C 1 atmospheric correction, 0 °C 1 atmospheric correction, and actual flow rate. SAMP. DATA STD SET 25deg 1atm RTC SET DISPLAY TRBL SET SERIAL No.
Page 27
Sampling Data In the main screen, press [SAMP. DATA] to review ■SAMPLING DATA DISPLAY previous sampling data (up to the last 5 sets). Press [1] to review the most recent sampling data. The most recent 5 sets can be viewed by pressing the other numbers respectively.
Page 28
The sampling data is shown over a total of 4 pages. Press [] and [] to toggle between the pages. SAMP No.: 1 Sampling start START time 0 9 / 0 9 / 0 3 1 4 : 0 0 : 0 0 SAMPLING Sampling time TIME...
Page 29
Detailed Error Display [TRBL] is displayed when an error has occurred. SAMP No.: 1 TRBL ESC START 0 9 / 0 9 / 0 3 1 4 : 0 0 : 0 0 SAMPLING TIME 8:00:00 VOL. 480.0m ▼ ▲ Press [TRBL] to view detailed error information.
Page 30
The screen will appear as in the figure, but only if a blower error occurs. BLOWER TRBL 「RPM」:Rotational error 「ERR」:Other error 0 9 / 0 9 / 0 3 1 4 : 3 0 : 0 0 Time of error occurrence * The blower will automatically stop if the internal temperature reaches or exceeds 95 °C.
Page 31
Setting the Screen In the menu screen, press [DISPLAY] to display the ■S e t t i n g M o d e screen setting screen in the figure at right. Press [+] and [-] to adjust the brightness. Press [ESC] to confirm the settings and return to the menu screen.
Page 32
Setting the Error In the menu screen, press [TRBL SET] to display the error setting screen as in the figure below. Press the respective numerical regions to proceed to the numerical value entry screen. Set the respective threshold values. Press [ESC] to return to the menu screen. [ESC] key Flow rate accuracy threshold Suction pressure upper limit...
Page 33
Calibration Mode In the menu screen, press [CAL] to switch to calibration mode. This section only describes the operating methods in the calibration mode. When performing an actual calibration, see the section on calibration methods. (See page 33.) [ZERO] key [SPAN] key Flow rate FLOW...
Page 34
Leak Check Press [LEAK CHECK] to switch to leak check mode. 0 7 / 2 9 1 0 : 2 3 : 4 5 25deg LEAK CHECK L/min For the leak check, prepare an HV circular rubber plate (sold separately). Attach it to the same procedures as for attaching filters, and seal off the sampler's suction port.
Calibration Methods Calibration is recommended to further improve the accuracy of sampling. 1 Point and 2 Point Calibration 1 point calibration is an effective method for adjusting the accuracy of a given value. A disadvantage however is that accuracy will actually be worsened if that value changes significantly. 2 point calibration can be used to adjust accuracy across a calibration range.
An optionally available standard flow meter can be attached to the sampler's intake side. Calibration at the work site is simplified by using Sibata OFD-1 orifice flow meter (sold separately) as the standard flow meter (Optionally available round/rectangular adapter for orifice flow meter is required.
Sampling data can be retrieved by connecting a PC using the USB port provided on the product. No particular communications software is prepared, but products such as HyperTerminal can be used. A special USB driver is required for communications. The USB driver can be downloaded from Sibata website: (http://www.sibata.co.jp/technology/technology-27640/)
1. One year has elapsed since flow rate calibration by the manufacturer. 2. The internal battery run-down icon is lit. 3. The packing or rubber parts have become stiff, or cracks have developed. If any other problems arise, contact your Sibata agent. - 36 -...
The internal battery is dead. The battery must be replaced. the clock reads "2000/1/1" (The internal battery icon is lit.) Contact your Sibata agent. when the power is turned OFF. There is no sampling data. The internal battery is dead.
Main Specifications Item Code 090130-31 Model HV-500R Standard Suction Flow Rate 500 L/min. (with glass fiber filter φ 110mm) Configurable Flow Rate Range 100 to 800 L/min Accuracy of Constant Flow Rate Within ±5 % of the set flow rate Suction Pressure *1 -160 hPa (500 L/min) Flow Rate Detection...
The unit can be attached to a particle size selector (option) to cut and sample particles over specified sizes and sample them. Please refer to an operation manual about details and the handling methods. Please contact your Sibata agent to find out how the unit should be combined depending on purpose of your use. Item code...
And, the Rectangular Filter Holder enables a sampling with 8" x 10" filter. Please refer to an operation manual of the Shuttle Tube for the handling and details. Please contact your Sibata agent to find out how the unit should be combined depending on purpose of your use...
Orifice adaptor For particle size selector, round type ※ Please contact your Sibata agent to find out how the unit should be combined depending on purpose of your use. ※1 When using this product, "080130-07511 Round adapter for orifice flow meter" or "080130-0551 Rectangular adapter for orifice flow meter"...
For details of repair after the Warranty has expired, contact your Sibata agent. The product shall be repaired for a fee only if Sibata judges that repair shall restore its functions, and its functions can be sustained in the future only in accordance with specified methods of use.
The sampler body is made almost entirely from metals (aluminum and stainless steel). Inquiries If you have any questions about this product, or if there is any other way in which we can be of assistance, contact your Sibata agent. 2017.08.29H (07) - 43 -...