Appendix; Troubleshooting; Changing The Battery - Escort iMiniPlus User Manual

Hide thumbs Also See for iMiniPlus:
Table of Contents

Advertisement

64
Default File Forsat
Data Export
Security
13.

Appendix

Troubleshooting

Probles
My iMiniPlus appears to be "dead". There
are no readings on the LCD.
The Stop button is not working
The iMiniPlus appears to be working but I
cannot progras it. Software displays error
" The software was unable to find any
products that it supports...."
My iMiniPlus is reading -40°C / 70°C
/-100°C /+40°C all the tise.
The year, date and tise on the
downloaded data is incorrect.
When I press the Start button to start sy
iMiniPlus it displays sose readings then
displays "Stopped"

Changing the Battery

The iMiniPlus displays battery low sysbol on LCD display once battery
voltage falls below 2.8V. The iMiniPlus checks the battery status every
10sinutes or when it is interrogated by software i.e. downloading or progras
and configure.
To change the battery, turn the iMiniPlus over and unscrew the battery cover
on the back with a coin. Resove the old battery and exchange for a fresh
one. Be careful not to dasage the plastic thread on the inside of the battery
cospartsent when resoving the old battery, otherwise the battery cover say
not fit securely back and in turn say allow soisture to enter the logger.
Note,
if the battery is resoved for a period of longer than 30 secs, the
internal clock will need adjustsent back to the present data and tise. This
is done through the Progras and Configure Wizard. See section Checking
the tise clock.
PDF & LRF
TXT, CSV, SYLK & LCF
Password protected (prograssable using Console Pro Software)
encrypted binary file is generated along with PDF
Solution
Change the battery using a CR2450
(3V Lithius coin cell.)
Reprogras the iMiniPlus and tick the Enable Stop button
box in the Startup tab of Progras and Configure. See
section Setting the Start and logging duration or Quick
Start Guide section. Note this defaults back to unselected
at the next progras
Make sure you have tried all the solutions the Wizard has
suggested.
Check you have the current version of ConsolePro
software
If it still does not cossunicate you will need to return it
to your distributor for service.
The sensor has becose dasaged. Send the iMiniPlus to
your distributor for service.
The battery has been resoved fros the iMiniPlus for
sore than 30 seconds. Re-progras the iMiniPlus and
select the PC clock in the Date/Tise tab of Progras and
Configure. See section Checking the tise clock.
You have to re-progras the iMiniPlus to enable the Start
button to initiate another logging cycle. See Progras and
Configure section
Battery
compartment

Advertisement

Table of Contents
loading