INSTALLATION & SERVICE INSTRUCTIONS
Unit Filter Installation
Bench Top Replaceable Filter (BTR)
Sample line string below:
FFU-HE/24/48/BTR/BC/ECM/115/10//CT//HEPA//ECMDX//WK/DSW/PC/TMB///ML/FL/WPF/PL-AL///PL-MILL//
NOTE:
Extreme caution should be taken to avoid contact with filter media. Touching filter media may result in filter failure.
Damage to filters can occur during installation or during leak qualification testing. The filters supplied by Price have been 100%
tested and certified by the manufacturer to be free of defects and leaks. Price cannot accept responsibility for damage that occurs
after shipment, whether through transit, handling or installation, and will not replace filters under our standard warranty.
NOTE:
If ordered with Filter, or Motor LED Indicator (ML/FL), refer to LED installation instructions which must be done at same time
as filter installation.
FFU-HE/24/48/BTR/BC/ECM/115/10//CT//HEPA//ECMDX//WK/DSW/PC/TMB///ML/FL/WPF/PL-AL///PL-MILL//
STEP 1
Place filter box on floor, careful to set box in correct orientation,
as per the arrow.
STEP 3
Place the filter gently on the floor, remove the tape, and take
the filter out of the bag.
STEP 2
Remove filter from box being careful not to touch filter media.
STEP 4
Place filter on a flat, clean, protecting surface with the
expanded face screen on the bottom. Make sure to protect
face LED if this has been selected as an option.