Problem
Frost occurs on head or
knuckle.
Frost occurs evenly over
the cylinder surface.
Miscellaneous frost spots
on cylinders.
Delivery gas is too cold.
In liquid delivery, liquid is
mixed with high amount
of gas.
In CO2 service, cylinder
does not deliver product
properly.
For further information contact Chart's Technical Service Department at (800) 400-4683.
Thermo Scientific
Problem Cause
Residual frost remains from
last fill or recent product
use.
Sight gauge is leaking.
The gas withdrawal rate is
high. The gas use vaporizer
is frosted.
Cylinder has lost vacuum.
Cylinder may have internal
damage.
Delivery rate exceeds rec-
ommended delivery.
Cylinder pressure is higher
than optimum for liquid with-
drawal.
Possible dry ice blocks have
formed in system.
Corrective Action
This is normal. Ice may remain for days after a
fill or heavy use.
Check for gas escaping from under sight gauge.
Refer to Service & Maintenance section for repair.
This is normal.
This is accompanied by high rate of pressure rise
or high loss rate. Call MVE for return instructions.
Call MVE for evaluation or repair/return information.
Refer to Operation section for recommended
maximum delivery rates.
Refer to Service and Maintenance section to reset
the cylinder pressure for liquid use. Also, use
a phase separator on the end of the transfer hose.
Refer to Service & Maintenance section for reliquefying
procedures.
Section 6
Troubleshooting
230 liter LN2 Supply Tank
6-5