Summary of Contents for Zeiss HBO 50

  • Page 2 In the case of technical problems or when service is required, please get in touch with your local Carl Zeiss representative or contact us at the address given below. Provided operating manuals: Power supply for HBO 50: ebq 52 ac lamp power supply for HBO 50 W Issued by: Carl Zeiss Microscopy GmbH Carl-Zeiss-Promenade 10 07745 Jena, Germany microscopy@zeiss.com...
  • Page 3: Table Of Contents

    Illustrations Fig. 1 Components of HBO 50 Microscope Illuminator ..............8 Fig. 2 Line spectrum of HBO 50 mercury-vapor short-arc lamp ........... 10 Fig. 3 Replacing the HBO 50 burner ................... 11 Fig. 4 ebq 50 ac lamp power supply ................... 13 Fig.
  • Page 4: Safety Notes

    Before any intervention in the device, manufacturers, (e.g. OSRAM). disconnect the power cable. CAUTION! Non-compliance with this safety note may result in damage to the equipment.  NOTE! Note to be observed in using the HBO 50 Microscope Illuminator. M60-2-0014 e 05/2012...
  • Page 5: Note On Illuminator

    0000-000) and protective gloves with artery compatibility protection (Cat. No. 417008-0000-000) when you remove the HBO 50/AC L1 / L2 burner from The HBO 50 Microscope Illuminator has been its packaging or when you deposit it there for produced and tested in compliance with disposal (e.g.
  • Page 6: Notes On Use

    Make sure to switch off the illuminator and keep a cool-down period of at least 20 minutes The light of the HBO 50 short-arc illuminator is before taking off the illuminator. Hold the of very high intensity and contains high portions illuminator on its cover, not on its cooling fins.
  • Page 7: Notes On Care And Maintenance

    Carl Zeiss Service. Notes on disposal − Replace the burner (lamp) after the mean The HBO 50 AC (L1 / L2) burners contain little lifetime quantities of mercury! (refer to data sheet of lamp manufacturer) has elapsed;...
  • Page 8: Description

    Microscope Illuminator Carl Zeiss Intended Use, Components HBO 50 Description The HBO 50 Microscope Illuminator consists of the following components (Fig. 1): − Lamp housing (1/1) with mount (1/6) (for Intended use connection to the microscope) and removable lamp compartment The HBO 50 Microscope Illuminator is designed −...
  • Page 9: Technical Data

    Burner Mercury-vapor short-arc lamp For technical data of burners, OSRAM refer to the specifications of the Type HBO 50 W/AC / L1 or HBO 50 W/AC / L2 manufacturers. acc. to Cat. No. 381619-0000-000 Recommended for use: Lamp voltage L1: 39 ... 45 V / L2: 34 ... 39 V Lamp current L1: 1.30 A / L2: 1.45 A...
  • Page 10: Fig. 2 Line Spectrum Of Hbo 50 Mercury-Vapor Short-Arc Lamp

    Description Microscope Illuminator Carl Zeiss Technical Data HBO 50 Relative Intensity / % Wavelength / nm Fig. 2 Line spectrum of HBO 50 mercury-vapor short-arc lamp M60-2-0014 e 05/2012...
  • Page 11: Start-Up

    3 Burner SW 3. 4 Lamp mount 5 Spring lever (lamp mount) • Press unlatching key (3/2) of HBO 50 down. 6 Spring lever ( cooling element) Remove lamp compartment completely 7 Lock screw of compartment from lamp housing (3/1) and put it Fig.
  • Page 12 Start-up Illuminator Carl Zeiss Inserting and replacing the burner HBO 50 • Press spring lever (3/6) of cooling element up and pull cooling element of transport safety device or of used burner (3/3) up and put it cautiously on the table.
  • Page 13: Starting Up The Ebq 50 Ac Lamp Power Supply

    Starting up the ebq 50 ac lamp power supply When the device is connected to the mains, Safe and reliable operation of the HBO 50 the full operating or starting voltage is available Microscope Illuminator requires that you know at its output (danger to life!).
  • Page 14: Focusing And Centering The Burner

    Carl Zeiss Focusing and centering the lamp HBO 50 Focusing and centering the The HBO 50 microscope illuminator is to be adjusted in two steps: coarse adjustment burner (without adjusting aid) and subsequent fine adjustment by means of the adjusting aid on the microscope.
  • Page 15: Fig. 7 Cathode Spots After Adjustment

     Adjustment with adjusting aid – Make sure that the two cathode spots • Attach the HBO 50 illuminator to the of the HBO 50 illuminator do not microscope stand following the description completely overlap in the centering...
  • Page 16: Removal Of Faults

    Replace the lamp by a new one (refer to Section Inserting and replacing the lamp). The two following measures may only be performed by authorized persons or Carl Zeiss Service: Check of the safety circuit of the illuminator. Replacement of starter.