Chapter 8 Setting of Scale's Linearity Adjustment

1. Operation Method:
▲!Before this operation, the calibration plug should plug in the 15 pin RS232 jack in the
back of the indicator.
Before revising, load the weight which needs to be revised (For example: a scale fits to
the linearity requirements on weight 0-30T but not on 40T, then the scale's revising
load point should be 40T)
Detailed operations are as follows:
Steps
Operation
Load
1、Ready
weight
revising point
2. Turn on
Press[Linearity
the
Correction]
indicator
Enter
and choose
[1][1][1][1][1][1
operating
],
mode
Press[Enter]
Press[0][1]
Press[Enter]
3、Enter
Enter
Revising
[4][0][0][0][0]
Press[Enter]
www.kelichina.com
Display
the
[******]
of
[-------]
[111111]
[S 00]
[S 01]
[000000]
[040000]
[040000]
Notes
When the scale has loaded revising point
weight
value,
the
exceeded or lacked weight value;
Ask to enter password
Only when the password is 111111, can the
next step continue, otherwise back to the
weighing state;
Default is 00 and do not carry on linearity
revising. If the number is bigger than 00
then carry on linearity revising.
Enter 01 and begin linearity revising;
Ask to enter revising point weight value.
Enter weight value to 40000kg
The indicator return to the weighing state
and display weight 40000kg;
indicator
displays
Page 25 of 38