Chapter 5 Corner Adjustment; Manual Corner Adjustment; Automatic Corner Adjustment - KeLi D2002E User Manual

Digital load cell and indicator
Table of Contents

Advertisement

1. Manual Corner adjustment

Please plug calibration plug into 15 pins RS232 at the back of indicator for manual Corner
adjustment. Please refer to the below chart for detailed information. (* original value, load
cell number starts 01, if there are 4 load cells in total, they are 01,02,03,04 )
Step
Operation
Press
Corner-Rectify]
1. Select the
number of load
Press [0][2]
cell
Press [Enter]
Press [1]、[0]、
[0]、[4]、[5]、
2Adjustment
[0]
press [enter]
Notice
please plug off calibration plug when operation finishes
When mistake happen which be caused by hot plug in or off, please press key "check" or
"power off" for a new set up

2. Automatic Corner adjustment

Generally speaking, digital load cell has a good sensitivity consistency. When the scale has been
correctly installed, it only needs slight hand adjustment according to the user's manual, or even do
not need adjustment. If it can not be adjusted equal by hand please apply to the auto adjustment.
Operating Steps:
Power off——Plug in the calibration plug——Power on
www.kelichina.com

Chapter 5 Corner adjustment

Display
[Manua
[no- 01]
[no- 02]
[1.00000]
[1.00450]
Remark
Plug in calibration plug under weighing
state
Default adjust load cell 01
command: check cell 02 ISN
Original Corner data
Adjust data into 1.00450(data is available
from 0.5000 to 2.0000)
When finish, it returns to weighing state