ProMOTION EV1
MENU OVERVIEW
History - Selecting 'History' gives access to the last 10 previously completed
programs. Note that the ProMOTION EV1 App gives further access - up to 100
programs:
1
Pressing 'Up' or 'Down' allows the user to view the program details.
2
Press 'Select' proceed to the summary screen.
3
Press 'Select' to initiate the program.
Programs - The 'Programs' function gives access to pre-programmed therapies.
10 slots are available, the first 3 of which are default programs created by
ProMOTION (see next page). Programs created in the ProMOTION EV1 App can
be saved in the remaining slots:
1
Select 'Programs', then press 'Up' or 'Down' to choose a program.
2
Press 'Select' to view a program's details. Contrast therapies will display
each cycle as a number e.g. 01/05, with the details of each cycle displayed
to the left of the screen.
3
Press 'Select' proceed to the summary screen.
4
Press 'Select' to initiate the program.
P30
User Manual
ProMOTION EV1
Default Programs
1
Cold: 6°C, 75mmHg for 20 minutes.
2
Hot: 40°C, 30mmHg for 20 minutes.
3
Contrast: Cycles between 6°C @ 75mmHg and 40°C @ 30mmHg
for 20 minutes
Precautions
Please do not remove components from each other while the device
is in use. Doing so may cause inadvertant system error.
If the Pump connection cable or Pump Unit are disconnected at any
stage, simply reconnect it. The Pump Unit will take a few moments to
recalibrate before continuing the program.
The Therapy Unit will not recall incomplete or interrupted
programmes. If the battery is removed during a program, the
device will reset from startup.
Please note that using the system in an environment with a high or low
ambient pressure or temperature may affect the ability for the unit to
provide adequate cooling, heating or compression.
Excessive movement during a program is not advised, as this can alter
pressure in the compression system, causing ineffective treatment.
Failure to properly follow the set-up instructions may result in
inadequate therapy. Ensure all components are present before use.
Power the Therapy Unit down before removing it from the cradle.
It is recommended that the ProMOTION EV1 device is disassembled
and placed back into protective casing for storage transport.
P31
User Manual