8.5.10 Parameter 12: Display Intercept "a" (Decimal) 000...999 [mg/m3]
To change the decimal part of the intercept a, follow the same procedure as for Parameter 11
8.5.11 Parameter 13: Display Slope "b" (Integer) 000...999 [mg/m3/mA]
To change the integer part of the slope b, follow the same procedure as for Parameter 11
8.5.12 Parameter 14: Display Slope "b" (Decimal) 000...999 [mg/m3/mA]
To change the decimal part of the slope b, follow the same procedure as for Parameter 11
8.5.13 Parameter 15: 20 mA scaling in 000...999 [IEU] x 1 000 000
This parameter is used for a very fine and accurate scaling and correlation of the 20 mA scaling to the
Inductive Electrification Unit (IEU).
IEU is described under chapter Error! Reference source not found. Error! Reference source not found.:
The reading of the IEU value at the Parameters 15 and 16 is the base for the Range setting on
the Sintrol Signal Generator. The Signal Generator is used for the linearity checks in regulated
applications.
As the IEU is a very fine scale with an open scale from 0 to several millions, the reading of this number is split
into two parameters.
Parameter 15: 20 mA scaling in 000...999 [IEU] x 1 000 000
Parameter 16: 20 mA scaling in 000...999 [IEU] x 1 000
To change the [IEU] x 1 000 000 part of the number, follow the below procedure:
Navigate to Parameter 15 by pressing K1
Key the desired value into the instrument.
Safe the result by pressing K3
Rescaling of the mA scale with the Parameters 15 and 16 will overwrite previous settings done
by using Parameter3 and vice versa.
8.5.14 Parameter 16: 20 mA scaling in 000...999 [IEU] x 1 000
To change the [IEU] x 1 000 part of the number, follow the same procedure as for Parameter 15

8.5.15 Info: Firmware Version

This is a read only Parameter and returns the Firmware Version.