8.5.8 Parameter 10: Command parameter

The command parameter has 3 functions:
001 Reset to factory defaults
002 Enable mg/m
003 Disable mg/m
To Reset to factory defaults, follow the below procedure:
Navigate to Parameter 10 by pressing K1
Key 001 into the instrument.
Safe the result by pressing K3
Now all parameters are set back to factory settings
This affects all parameters including the result of the Auto setup procedure. All previously
done settings will be over written.
To show on the local display the measurement in mg/m
Enter the definition of the calibration function yi = a + b xi by changing the parameters 1., 2., 3., and 4.
Navigate to Parameter 10 by pressing K1
Key 002 into the instrument.
Safe the result by pressing K3
Navigate to Parameter 1 by pressing K1
Key 002 into the instrument.
Safe the result by pressing K3
Now the measurement is displayed in mg/m
To disable the measurement in mg/m
Navigate to Parameter 10 by pressing K1
Key 003 into the instrument.
Safe the result by pressing K3
Now the measurement will not anymore be displayed in mg/m
calibration
3
calibration
3
3
3
, follow the below procedure:
3
, follow the below procedure:
3