Logic Test; Operation - Jeulin CL 3020 Manual

Table of Contents

Advertisement

Available languages

Available languages

Measurement and instrumentation
CL 3020 multimeter
Ref : 291 175
ENGLISH

2.11 Logic Test

Threshold:
o
high logic level (2.8 ± 0.8 V)
o
Low logic level (0.8 ± 0.5 V)
Indication:
o
40 ms buzzer at low logic level.
Frequency response:
o
20 MHz.
Detectable pulse width:
o
25 ns.
Pulse limits:
o
rate > 30 % & < 70 %.
Overload protection:
o
500 V DC or AC RMS.
2.12 Temperature
Ranges:
–20°C to +750°C.
o
Resolution:
o
1°C.
Accuracy:
o
± (2.0 % of the displayed value +3 digits) from -20°C to
+500°C
o
± (3.0 % of the displayed value +2 digits) from +500°C to
+750°C.

3 Operation

Read carefully the information relating to safety before carrying out
measurements. Check the device to detect any damage, contamination
(excessive dust, grease, etc.) and possible defects. Examine the test cords
to check if the insulation is not damaged or the wires are not cut off or
shredded. If abnormal conditions are detected, do not perform the
measurement.