Table of Contents

Advertisement

Available languages

Available languages

Measurement and instrumentation
CL 3020 multimeter
Ref : 291 175
ENGLISH
2.6 Continuity
Audible signal: less than 40  ± 20 
Overload protection: 500 V DC or AC RMS

2.7 Diode test

Test current: 1.0 mA  0.6 mA
Accuracy: ± (3.0% of the displayed value +3 digits)
Open circuit voltage: 3.0 V DC typical
Overload protection: 500 V DC or AC RMS.
2.8 Capacitance
Ranges of values:
o
4 nF, 40 nF, 400 nF, 4 µF, 400 µF
Accuracy:
o
± (5.0% of the displayed value +10 digits) for all the other
ranges < 100 µF
o
± (8.0% of the displayed value + 10 digits) above 100 µF.
2.9 Inductance
Ranges:
o
4 mH, 40 mH, 400 mH, 4 H, 40 H.
Accuracy:
o
± (5.0% of the displayed value +20 digits) for the 4 mH range
o
± (5.0% of the displayed value +10 digits) for all the other
ranges.
Test frequency:
Ranges 4 mH and 40 mH - 1 KHz
Ranges 400 mH and 4 H - 270 Hz
Range 40 H: 27 Hz
Test conditions: Over-voltage ratio > 5 to 270 Hz
2.10 Frequency (Automatic range change)
Ranges of values: 4 kHz, 40 kHz, 400 kHz, 4000 kHz.
Accuracy: ± (0.5 % of the displayed value +1 digit) for all the other
ranges.
Sensitivity: Minimum 1.0 V RMS.
Overload protection: 500 V DC or AC RMS.