No Serial Connection With Com Port (Line Doesn't Rotate) - IGA ODP 100 User Manual

Table of Contents

Advertisement

IGA Manufactures Association
3.2

No Serial Connection With COM Port (Line doesn't rotate).

Check System Manager for an installed USB Serial Port.
With the windows Hyper Terminal you may check the serial port.
By typing in a capital "V", you'll get the revision description from the ODP 100.
By typing in various small letters, the ODP 100 responds with "?" .
If a connection with the Hyper Terminal does not work, the ODP 100 will not be able
to generate a connection either.
You may find certain USB Ports which are incapable to handle a USB Serial Port.
Take following actions to start USB Serial Port:
a.)
Switch Printer Off And On.
b.)
Remove USB Cable From The USB Connector. Insert USB Cable Into The
USB Connector.
c.)
Unplug Power Cable. Plug Power Cable Back In.
d.)
Shut down and restart ODP Manager.
e.)
If the USB Serial Port is not recognised by the operation system remove driver
and install driver once again.
10/2004
User Manual ODP 100
33

Advertisement

Table of Contents
loading

Table of Contents