Thermo Scientific
Warnings and Limitations - Latches (continued)
5. If the spring pin breaks, the handle becomes inoperative. Replacement of the
spring pin with one not identical to the original may cause a hazardous
condition by destroying the independence of the cam and handle.
Additionally, the bearing pin, handle is individually fitted.
6. Because of the rather large tolerances involved in castings, each latch-strike
combination must be individually aligned and alignment must be rechecked
whenever latches are replaced to avoid possible malfunctions.
7. In the event of a maximum violent explosion, the front door will open
rapidly and with little or no warning. It is therefore recommended that the
area in the arc of the door be marked as a danger area, perhaps by "red
striping".
General Information Section 2
Freas Safety Oven
2-3