MD 11
Motor type 21, 40,000 rpm
Motor type 21, 40,000 rpm/Min. speed
Motor type 21, 40,000 rpm/Max. speed 5000 – 40,000 rpm
Motor type 21, 40,000 rpm /Start ramp
Motor type 21, 40,000 rpm /Stop ramp
The maximum torque of the motor is boosted while starting the reverse rotation to disengage stuck
implants or screws.
Reverse torque
Reverse torque/Increase
Reverse torque/Increase time
Attention: Upon resetting to factory default all parameters will appear with factory default values
(except date and time).
Resetting to
factory default
Default value/Set default value
Exit from Configuration Menu
• To leave the Configuration Menu press "HP" and "PR" for 3 sec., until you hear a long tone.
MD 11, V2.0, Operating Instructions, REF 31665, V22/17
Range
Factory
default
300 – 5000 rpm
300 rpm
40,000 rpm Set max. speed of the motor.
1 – 1000 ms/10,000 rpm 100 ms
1 – 1000 ms/10,000 rpm 50 ms
Range
Factory
default
5 – 30 %
25 %
100 – 2000 ms
500 ms
Choice
Factory
default
Yes/No
Definition
Set the min. speed of the motor.
Set acceleration time to 10,000 rpm
Set breaking time from 10,000 – 0 rpm
Definition
Increase of the selected torque in reverse
rotation.
Time during which torque is raised.
Definition
Resetting all parameters of the Configuration
Menu to factory default.
EN
19