Page 2
U.S. and foreign patents and/or pending � patents. TEKTRONIX, TEK, SCOPE-MOBILE, and registered trademarks of Tektronix, Inc. TELEQUI P MENT is a registered trademark of Tektronix U.K. Limited. Printed U.S.A. Specification price change privileges are reserved.
(cont) Introduction Performance Characteristic The Tektroni x P6021 5 foot and 9 foot Current Probes are designed to measure alternatin g current waveform s. These probes, used with current devices such as the Type Risetime (with 580 134 Current Probe Ampl ifier, 7 A 14 Current Probe Ampl i...
200 mA and termination contain adjustments to optimize perform ance, matching the input characteristics of the vertical am plifier. To use the probe, first connect the P6021 Termination to the vertical input of the oscilloscope. Then connect the P602 1 Current Probe to the termination (see Fig. 2-1 ).
With a strong external field, the external field interference may the P6021 Termination switch set to 2 mA/mY and the os be minimized by the use of two current probes and a dif...
As indicated on the probe body, the turns ratio of the Introduction P6021 Curent Probe is 125: 1 . This refers to the number of windings in the secondary of the probe transformer. This section of the manual contains descriptions of the two turns ratios manufactured by Tektronix, Inc.
This section provides procedures to be used in checking ating within the given specifications. If equipment is substi the performance or in calibrating the P6021 Current Probe tuted, it must meet or exceed the specifications of the rec and Termination.
Page 12
Scale ilium As desired Volts/Cm General Variable Volts/Cm Gain The following procedure is arranged so that the P6021 Current Probe and Termination can be calibrated with the Input Coupling least, interaction of adju ents and reconnection of equip T rigger Slope...
Page 13
Termination to the input of the test oscilloscope. Set the termination slide switch to the 2 mA/mY position. b. Connect the P6021 Probe from the Calibration Fixture to the termination. c. Adjust the pulse generator output amplitude for four centimeters of deflection.
Performance Check! Calibration-P6021 Check Sensitivity a. Remove the P6021 Probe, Termination, and Calibration Fixture from the test equipment. b. Connect two 50-ohm GR lOX attenuators and a five nanosecond length of RG 213/U cable to the output of the 'pulse generator.