Quick Start:

1. Turn off the power.
2. Plug in the proper transducer- Turn on-wait for the
display to show zero (do not apply torque during this time).
The meter reads the transducer smart chip when turned on.
3. Apply torque, (see warnings) you will see the reading
increase as you increase the torque.
4. Clear the display with the clear key.
Or
5. Save torque to memory with the save key.
DO NOT OVER TORQUE TRANSDUCER!

To change the mode of operation

1. Press the mode key until "peak", "track" or "Fpeak"
shows.
2. Press the save key to accept the new mode.
Track mode displays the torque as it increases and
decreases.
Peak mode tracks up but will hold on the peak value.
Fpeak (first peak) mode will capture the click point of a
micrometer torque wrench and disregard further input.
This mode is used to test and calibrate click type torque
wrenches.

To change the units of measure

1. Press the units key until the arrow points to the desired
units (the set-up icon will also be on).
(EEEE will show if units are out of range for the transducer
attached).
2. Press the SAVE key to select the new units of measure.
Note: When you change the units of measure the limits
values (target or high and low) will be re-set to zero. The
meter will now power-on with the new units automatically.
1. Press limits key once to turn on.
2. Press the limits key again to turn off.
Note:
You will see the limits value in the secondary set of digits-
"%" and the current target torque will show if % is the
default or high and low arrows with the actual high and low
values if the high and low method of limits is selected (see
General Operation

To turn limits on/off