Troubleshooting:

Meter reads "OPEN" when turned on: 1. Transducer is not plugged in.
Meter won't show zero:
Meter beeps when pressing save button:
Degraded battery life:
Plug in transducer
2. There is a significant load on the transducer
when the meter is powered on.
Remove load (such as hanging wrench).
3. The transducer or cable may have been damaged.
Return for repair.
1. There may be a load on the transducer when power
was turned on:
Remove load and press Zero Tare or
Remove load and turn meter off then on
1. Memory is full.
Clear Memory
1. Recharbeable battery has reached re-charge limit:
Replace rechargeable 9 volt battery.
Use conventional Lion battery (IMPORTANT!
Do not use the power adapter when running
the meter with a conventional (non-rechargeable)
battery.