Preventive Maintenance & Security Controls - NAUSICAA MAXXI UP 300 Manual

Table of Contents

Advertisement

Preventive Maintenance & Security Controls
Stand-Up Lift: MAXXI UP 300
Serial Number :
Control of security points (Frequency: depending on use, recommended every three months.
1
Connection head of the actuator (bolted)
2
Connection base of the actuator (bolted)
3
Connection head of the mast (bolted)
4
Locking base of the mast
5
Fixing of the castors (bolted)
6
Sliding of the sub-patellar support (adjustable knob)
7
State and sticking of the anti-slip material
8
Fixing of the mast handle
9
Overall state of the device (welds, joints, oxidation)
Control of electrical functions (Frequency: depending on use, at least once a year) this control
A
Operation of the actuator
B
Operation of the emergency system
C
Batteries status (charge holding)
D
Handcontrol status
Inspection carried out on :
4
Manufactured by NAUSICAA Médical S.A. / Approved by Ghizlane Labrosse (Biomedical Engineer)
This is a sight control : any wear trace must lead to a replacement.)
must be carried out when the hoist is loaded.
Inspection carried out by :
User Manual - Stand-Up Lift: MAXXI UP 300
GOOD
WRONG
GOOD
WRONG
Date of the next inspection :
DATE OF CHANGE
DATE OF CHANGE