Problem
The product, without patient, does
not go down.
Mobile parts are too hard, product is
difficult to handle.
The actuator does not work but the
user can hear a "click in the control
box when the remote control button
is pushed.
The actuator does not work and the
user does not hear a "click" in the
control box when the remote control
button is pushed.
The actuator stops when lifting.
The actuator does not work at all
12
Manufactured by NAUSICAA Médical S.A. / Approved by Ghizlane Labrosse (Biomedical Engineer)
T roubleshooting Guide
Origin
Our products need some weight
to go down.
Mobile parts need to be greased.
1 . Batteries are unloaded.
2 . The actuator jack is not plugged.
3 . The actuator jack is deteriorated.
4 . There are damages on the actuator or
on the control box.
1 . Emergency button is activated.
2 . There is no contact between the
battery and the control box.
3 . Batteries are unloaded.
4 . The remote control jack
is not plugged.
5 . The remote control jack is
deteriorated.
6 . There are damages on the remote
control or on the control box.
1 . Batteries are unloaded (sound system
+ "low battery" written on the screen).
2 . The product has been used
for too long.
3 . The weight lifted is too heavy.
1 . The control box is broken down.
2 . The remote control is broken down.
User Manual - Stand-Up Lift: MAXXI UP 300
Solutions
Push down on the lifting arm while
pushing the lowering button on the
remote control.
Grease the mobile parts.
1 . Load the device.
2 . Plug the actuator jack.
3 . The actuator has to be repaired.
4 . The entire electric system has
to be checked.
1 . T urn the button to unlock the device.
2 . Take off the battery and put it back.
3 . Check the charging cord.
4 . Plug the remote control.
5 . Change the remote control.
6 . The entire electric system has to be
checked.
1 . Check the charging cord then
load battery.
2 . Let the device rest a moment (as a
protection for the actuator lifespan).
3 . Reduce the weight lifted.
1 . The control box has to be changed.
2 . The remote control has to
be changed.