Motorization Technicals Characteristics
Cycle counter :
We have a time counter of 600,000 sec. If we consider that the actuator makes full strokes at an average speed of
9 mm/s that represents approximately 10,000 full cycles.
To check the number of cycles already performed, you must use the remote control 2 buttons:
-Press the 2 buttons at the same time.
The indication of the number-of cycles performed is indicated by the digits, the indication remains displayed for
about 3 seconds.
-If no digit is displayed the number of cycles performed is less than 150, 000 sec.
-If one digit appears the jack has been used for 150, 000 sec or 25% of the maximum time.
-If 2-digit appear, the actuator was used for 300, 000 sec or 50% of the maximum time.
-If 3 digits appear,the actuator was used for 450, 000 sec or 75% of the maximum time.
-If 4 digits appear,the actuator was used for 600, 000 sec or 100% of his time and the maximum box no longer work.
This counter work with time and not cycle because usual cycles counter count when you push on a remote control
button. The problem with this type of counter is when you push 5 times on one cycle, It'll count 5 cycles. The time
counter is more accurate and after 10 000 full cycles the actuator need to be checked by the manufacturer.
- After 10,000 complete cycles, a key appears on the screen and the actuator must be checked.
Overload protection system :
• If you attempt to lift a person weighing beyond the maximum limit of the device, an anvil with "kg" written will
appear asking you to immediately relieve the unit because it is a non-conform use.
Preservation system of the actuator :
• An integrated system for maintaining the actuator is included in the control box. It corresponds to a ratio between
the operating time and the rest time of the actuator.
• For 10% of usage time, it takes 90% of sleep time.
• For 6 minutes of continuous use, you must wait 54 minutes before you could reuse the device.
If this period is not respected, the device will not work and a jack will be displayed on the screen. When this cycle
is exceeded, there is a risk from overheating the equipment and cause a failure.
Soft stop and soft start Actuator Programming :
1
Push on the Emergency stop button
Push on button 1 and turn off the emergency stop button
The Control box Green led is flashing :
-3 times, Soft stop and soft start available
-2 times, Soft stop and soft start unavailable
Release button 1 on the handset after cessation of the green LED.
User Manual - Stand-Up Lift: MAXXI UP 300
11
Manufactured by NAUSICAA Médical S.A. / Approved by Ghizlane Labrosse (Biomedical Engineer)