Ventilator operating specifications

Ventilation operating modes
Volume Control
Pressure Control
Synchronized Intermittent Mandatory Ventilation (SIMV)
Pressure Support (PSVPro) with Apnea Backup ventilation —
(optional)
Non-adjustable
25.4 cm/10 in
91.5 cm/36 in
85°
Ventilator (V
) parameter ranges
T
Tidal volume range:
Incremental settings:
Minute volume range:
Pressure (P
) range:
Inspired
Pressure (P
) range:
limit
Pressure (P
) range:
support
Rate:
Inspiratory/expiratory ratio:
Inspiratory time:
Trigger window:
Flow trigger:
Inspiration termination level:
Backup mode delay:
20 to 1500 mL (Volume
Control and SIMV modes)
5 to 1500 mL (Pressure
Control Mode)
20 to 100 mL
(increments of 5 mL)
100 to 300 mL
(increments of 10 mL)
300 to 1000 mL
(increments of 25 mL)
1000 to 1500 mL
(increments of 50 mL)
0 to 99.9 L/min
5 to 60 cm H
O
2
(increments of 1 cm H
O)
2
12 to 100 cm H
O
2
(increments of 1 cm H
O)
2
Off, 2 to 40 cm H
O
2
(increments of 1 cm H
O)
2
4 to 100 breaths per
minute for Volume Control
and Pressure Control;
2 to 60 breaths per minute
for SIMV, PSVPro and
SIMV–PC+PSV (increments
of 1 breath per minute)
2:1 to 1:8 (increments of 0.5)
0.2 to 5.0 seconds
(increments of 0.1 seconds)
(SIMV and PSV Pro)
0 to 80% (increments of 5%)
0.2 to 1.0 L/min
(increments of 0.2 L/min)
1 to 10 L/min (increments
of 0.5 L/min)
5 to 75% (increments of 5%)
10 to 30 seconds
(increments of 5 seconds)