Datalogger Filling (Percentage); Dpc - BASI BFM921 User Manual

Electromagnetic flowmeter
Table of Contents

Advertisement

Electromagnetic Flowmeter BFM921
Auxiliary volume counter Reset
CLRAV
Command resets "Auxiliary volume counter".
BFM921 confirms execution with string „Ok<cr>".
Example:
Command „CLRAV<cr>" resets the Auxiliary volume counter.
Timed volume counter Reset
CLRTV
Command resets "Timed volume counter".
M921 confirms execution with string „Ok<cr>".
Example:
Command „CLRTV<cr>" resets the Timed volume counter.
5.2.2.6 Datalogger parameters
Datalogger step
DST(?)<CPD> { 0 | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 }
Datalogger can be set:
0
Datalogger is Off
1
Datalogger sampling rate is 5 minutes.
2
Datalogger sampling rate is 10 minutes.
3
Datalogger sampling rate is 15 minutes.
4
Datalogger sampling rate is 30 minutes.
5
Datalogger sampling rate is 45 minutes.
6
Datalogger sampling rate is 60 minutes.
7
Datalogger sampling rate is 120 minutes.
8
Datalogger sampling rate is 180 minutes.
9
Datalogger sampling rate is 240 minutes.
M921 confirms datalogger step with string „Ok<cr>".
Example RS232:
Command „DST0<cr>" switches datalogger "Off". If query „DST?<cr>" is sent, flowmeter returns response
in format „0<cr>".
Datalogger number of samples
DNR?
Response contains number of flowrate samples stored in datalogger.
Example:
If query „DNR?<cr>" is sent, flowmeter returns response in format „252<cr>" for 252 samples in datalogger.

Datalogger filling (percentage)

DPC?

Response contains datalogger filling in percent.
Example:
If query „DPC?<cr>" is sent, flowmeter returns response in format „14<cr>" for 14% datalogger full.
48