LDetek MultiDetek2 User Manual page 55

Gas chromatograph for multiple impurities
Table of Contents

Advertisement

Dev Alert:
The percentage of deviation allowed before the oven temperature
deviation alert
Low Limit:
The lowest value to be considered as a normal condition. If we have
reading lower than that, the system will stop trying to control it.
Because there is probably something wrong, a broken sensor for
instance.
Hi Limit:
The highest value to be considered as a normal condition. If we
have reading higher than that, the system will stop trying to control
it. Because there is probably something wrong, a broken cable for
instance.
Lookup Table:
This is the table to convert the sensor reading into units
Time Table:
This table contains the setpoint for the oven.
The "(at start)" is the setpoint to reach when no analysis is running.
In other words, before an analysis start. Then you can add more
setpoint at for different time in an analysis.
MULTIDETEK 2 User's manual
48