The user can change the SET button to any other convenient set point (see Settings). In
addition, the control knob is speed-sensitive; turning it faster produces a faster change in the
output.
The normal industrial process control loop current range, or span, is 4 mA to 20 mA (0% to
100%). Intermediate points are as follows:
Current
Percent
Settings
The IPM-100 has two selectable settings: Automatic Power-off and the current/percent
set point of the SET button.
To change Automatic Power-off setting:
1.
Press and hold the SET button while turning the unit on.
2.
After the firmware revision appears on the display, release the SET button. BAT and
ON indicate that Automatic Power-off is enabled.
3.
Press the MODE button to change the Automatic Power-off function (ON or OFF).
4.
Press the SET button to store your selection.
To change current/percent set point for SET button:
1.
Use the control knob to set the output to the desired value.
2.
Press the SET button to store the selected value.
8
0 mA
4 mA
–25%
0%
8 mA
12 mA
25%
50%
16 mA
20 mA
75%
100%
24 mA
125%