Gamma Adjust

By using the front sensor, Gamma Adjust can recalibrate the luminance and gamma. All PICTURE MODES are updated when
Gamma Adjust is performed.
1. To show the CALIBRATION menu, select the STAND ALONE CALIBRATION in the advanced menu (page 24).
2. Select GAMMA ADJUST in the MODE selection (Figure 1).
3. Select the target UNIFORMITY. Press SELECT to go to the next section.
4. Set the target LUMINANCE (cd/m
depending on user setting.
5. DICOM Measurement is continuously executed automatically.
After completing DICOM Measurement, you can set DATE and TIME using control keys.
2
) (Figure 2). Press SELECT to start the calibration. This may take several minutes
Figure 1
English-27
Figure 2