Agilent Technologies 6032A Service Manual page 51

Autoranging system dc power supply
Hide thumbs Also See for 6032A:
Table of Contents

Advertisement

(A8U6 = P/N 5080-2160 REV A.00.00, A.00.01 A.00.02, and A.00.04)
Inputs:
Node
U11-1
U11-2 U12-4,10
U16-10
Procedure:
Node
S1 LCL
S1 LCL
S2 OUTPUT ADJUST
S2 OUTPUT ADJUST
S3 DISPLAY OVP
S3 DISPLAY OVP
S4 DISPLAY SETTINGS
S4 DISPLAY SETTINGS
S5 FOLDBACK
S5 FOLDBACK
U12-5
U12-9
Set scope for dual trace operation, 2V/div, 10 ms/div, normal triggering, and positive edge on channel A.
Connect channel A to U12-3 and channel B to U12-2.
U12-3
Rotate RPG CW
U12-2
Rotate RPG CW
U12-3
Rotate RPG CCW
U12-2
Rotate RPG CCW
50
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Table 3-7. Front Panel RPG Latches and Input Port
Measurement
1ABC
Toggling (unstable)
9H84
released
U11-13 Lo
depressed
U11-13 Hi
released
U11-14 Lo
depressed
U11-14 Hi
released
U11-3 Lo
depressed
U11-3 Hi
released
U11-4 Lo
depressed
U11-4 Hi
released
U11-5 Lo
depressed
U11-5 Hi
Node toggles when RPG is rotated clockwise
Node toggles when RPG is rotated in either direction
U11-7, U16-9
Lo
U11-7, U16-8,9
Toggling
U11-7, U16-9
Lo
U11-7, U16-8,9
Toggling
U11-7, U16-9
Lo
U11-7, U16-8,9
Toggling
U11-7, U16-9
Lo
U11-7, U16-8,9
Toggling
U11-7, U16-9
Lo
U11-7, U16-8,9
Toggling
U16-8 Hi
U16-8 Hi
U16-8 Hi
U16-8 Hi
U16-8 Hi

Advertisement