CHANGE 12:
Under A1 Main Board,
*********** For Model 6031A only ***********
Change C23 and C28 to .047uF Agilent p/n 0160-
5895.
Add C21 and C24, .047uF Agilent p/n 0160-5895.
Add R21 and R22, 1 ohm p/n 0699-0208.
*********** For Model 6032A only ***********
Change C20 and C22 to .047uF Agilent p/n 0160-
5895.
Add C21 and C23, .047uF Agilent p/n 0160-5895.
Add R20 and R23, 1 ohm p/n 0699-0208.
CHANGE 13:
Under A8 GPIB Board, change U8 to p/n 1818-4111.
CHANGE 14:
Under A1 Main Board, change C11,12 to 0.047uF, p/n
0160-5895. On page 5-3, change R12 and R13 to 2.2
ohms, p/n 0699-0188.
CHANGE 15:
Under A8 GPIB Board, delete D22.
CHANGE 16:
Under A5 Diode Board, delete L3.
CHANGE 17:
Under A1 Main Board, change U1 to p/n 1906-0218.
CHANGE 18:
Under A6 AC Input Filter, change L1 to p/n 06011-
80094.