Agilent Technologies 6032A Service Manual page 17

Autoranging system dc power supply
Hide thumbs Also See for 6032A:
Table of Contents

Advertisement

g.
Adjust A8R51 (READBACK ZERO) until the value displayed on the controller toggles between:
0 and 50mV (6030A)
0 and 5mV (6031A)
0 and 15mV (6032A)
0 and 125mV (6035A)
h.
After adjusting A8R51 you must continue the calibration procedure through to the completion of Constant Voltage
Zero Calibration. Remember to disconnect the external power supply and resistor.
Figure 2-2. Remote Readback Zero And CV Zero Calibration Setup
16