Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic
emissions IEC
61000-3-2
Voltage
fluctuations/ flicker
emissions IEC
61000-3-3
Declaration – electromagnetic immunity
The Philips Avent breast pump is intended for use in the electromagnetic environment specified
below. The user of the Philips Avent breast pump should assure that it is used in such an environment.
IMMUNITY test
Electrostatic discharge
(ESD) IEC 61000-4-2
Electrical fast transient/
burst IEC 61000-4-4
SurgeIEC 61000-4-5
Compliance
Electromagnetic environment – guidance
Group 1
The Philips Avent breast pump uses RF energy only for its
internal function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby electronic
equipment.
Class B
Class A
The Philips Avent breast pump is suitable for use in all
establishments, including domestic establishments and those
directly connected to the public low voltage power supply
network that supplies buildings used for domestic purposes.
Complies
IEC 60601 test level Compliance level
± 6 kV contact± 8 kV air
± 2 kV for power supply
lines
± 1 kV line(s) to line(s)
± 6 kV contact± 8 kV air
± 2 kV for power supply
lines
± 1 kV line(s) to line(s)
Electromagnetic
environment –
guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered
with synthetic material,
the relative humidity
should be at least 30 %.
Mains power quality
should be that of a
typical residential or
hospital environment.
Mains power quality
should be that of a
typical residential or
hospital environment.