10. Power Line Surge Test (IEC 61000-4-5 Surge)

10.1

Climatic Condition

The climatic conditions shall be within the following ranges:
10.1.1
Ambient Temperature: 15 C to 35 C
10.1.2
Relative Humidity: 10% to 75%
10.1.3
Atmospheric Pressure 86kPa(860 mbar)to 106kPa (1060mbar)
Note: The temperature and relative humidity should be recorded in the test report.
10.2

Test Conditions:

10.2.1
Wave-shape of the current surge: (refer to IEC 61000-4-5)
Open circuit voltage:
Short circuit current:
10.2.2
Polarity: positive / negative
10.2.3
Phase shifting:in a range between 0º to 270º versus the AC line phase angle
10.2.4
Repetition rate:at least 1 per minute
10.2.5
Number of tests:
10.3
The surge will be applied between lines and between lines and ground.
10.4
If not otherwise specified, the surge to power supply circuits shall be applied synchronized to
the voltage phase at the zero crossing and the peak value of the AC voltage wave. (Positive
and negative)
10.5
The surge voltage for test is from 1 KV and increases 1 kV for each step.
10.6
The recommended severity levels for the surge voltage test is 2.0 KV, and without any
degradation or loss of function that is not recoverable due to damage of component or
software allowed.
10.7
Display set high-resolution mode, AC input use AC 240V
(1.2 / 50 µs)
(8 / 20 µs)
at least 5 positive and 5 negative at the selected points.