8 ELECTROMAGNETIC COMPATIBILITY

8.1 General

The HFG01 has been designed to meet the requirements of EU Directive
2014/30/EU. Compliance was demonstrated by testing to the requirements of the
following standards: -

8.2 Emissions

EN61326-1 (2013) consisting of
• EN55011 (2009) + A1 (2010) Radiated RF emissions (Enclosure) and
Conducted RF emissions (applies to the HFG01 supply adapter), Class A

8.3 Immunity

Enclosure:
EN61326-1 (2013) consisting of
• EN61000-4-2 (2009) Electrostatic Discharge immunity, 4 kV Air & Contact
Discharge.
• EN61000-4-3 (2006) + A1 (2008) + A2 (2010) Radiated RF immunity
(Enclosure), 3 V/m.
AC power:
These apply to the HFG01 supply adapter only:
EN61326-1 (2013) consisting of
• EN61000-4-11 (2004) Voltage dips/short interruptions
• EN61000-4-4 (2004) + A1 (2010) Fast Transient common mode (AC mains
port)
• EN61000-4-5 (2006) Surge (AC mains port)
• EN61000-4-6 (2009) Conducted RF immunity (AC mains port)
• EN61000-3-2 (2006) Harmonic emission (AC mains port)
• EN61000-3-3 (2013) Voltage fluctuation (AC mains port)

8.4 Non-Applicability

EN61326-1 (2013) consisting of
• EN61000-4-4 (2004) Fast Transient common mode (signal ports)
• EN61000-4-5 (2006) Surge (signal ports)
• EN61000-4-6 (2009) Conducted RF immunity (signal ports)
These phenomena are deemed non applicable on the basis that the HFG01 has
no relevant signal ports. The AC test port of the HFG01 may be looked upon as a
signal port but the captive lead is less than 3m in length hence the tests do not
apply.
The test port is not connected directly to the public low voltage supply network and
so tests relating to AC ports are only applicable to the AC/DC adapter used for the
HFG01 supply input.