Avtech product is found to be defective, Avtech shall at its option repair or replace said defective item. This warranty does not apply to units which have been dissembled, modified or subjected to conditions exceeding the applicable specifications or ratings.
TABLE OF CONTENTS WARRANTY........................2 TECHNICAL SUPPORT....................2 TABLE OF CONTENTS....................3 INTRODUCTION.......................5 SPECIFICATIONS......................6 REGULATORY NOTES....................7 FCC PART 18.......................... 7 EC DECLARATION OF CONFORMITY..................7 DIRECTIVE 2011/65/EU (RoHS).....................8 DIRECTIVE 2002/96/EC (WEEE)....................8 FIRMWARE LICENSING......................9 INSTALLATION......................10 VISUAL CHECK........................10 POWER RATINGS........................ 10 CONNECTION TO THE POWER SUPPLY................10 PROTECTION FROM ELECTRIC SHOCK................11 ENVIRONMENTAL CONDITIONS..................12 LABVIEW DRIVERS......................
4% (or 8% with the -XP1 option, or 16% with the -XP2 option). Rise and fall times are fixed at less than 10 ns. The AVR-3HF-B includes an internal trigger source, but it can also be triggered or gated by an external source. A front-panel pushbutton can also be used to trigger the instrument.
ISO/IEC 17025:2005 accredited calibration laboratory. However, Avtech itself does not claim any accreditation. For applications requiring traceable performance, use a calibrated measurement system rather than relying on the accuracy of the pulse generator settings.
If interference is observed, check that appropriate well-shielded cabling is used on the output connectors. Contact Avtech (info@avtechpulse.com) for advice if you are unsure of the most appropriate cabling. Also, check that your load is adequately shielded. It may be necessary to enclose the load in a metal enclosure.
< 1000 ppm (0.1% by mass) DIRECTIVE 2002/96/EC (WEEE) European customers who have purchased this equipment directly from Avtech will have completed a “WEEE Responsibility Agreement” form, accepting responsibility for WEEE compliance (as mandated in Directive 2002/96/EC of the European Union and local...
Article 9 of Directive 2002/96/EC. Customers who have purchased Avtech equipment through local representatives should consult with the representative to determine who has responsibility for WEEE compliance. Normally, such responsibilities with lie with the representative, unless other arrangements (under Article 9) have been made.
INSTALLATION VISUAL CHECK After unpacking the instrument, examine to ensure that it has not been damaged in shipment. Visually inspect all connectors, knobs, liquid crystal displays (LCDs), and the handles. Confirm that a power cord, a GPIB cable, and two instrumentation manuals (this manual and the “Programming Manual for -B Instruments”) are with the instrument.
5. Do not attempt any repairs on the instrument, beyond the fuse replacement procedures described in this manual. Contact Avtech technical support (see page 2 for contact information) if the instrument requires servicing. Service is to...
6. no pollution or only dry, non-conductive pollution. LABVIEW DRIVERS A LabVIEW driver for this instrument is available for download on the Avtech web site, at http://www.avtechpulse.com/labview. A copy is also available in National Instruments' Instrument Driver Library at http://www.natinst.com/.
FUSES This instrument contains four fuses. All are accessible from the rear-panel. Two protect the AC prime power input, and two protect the internal DC power supplies. The locations of the fuses on the rear panel are shown in the figure below: Fuses #1 and #2 Fuse #4 Fuse #3...
DC FUSE REPLACEMENT The DC fuses may be replaced by inserting the tip of a flat-head screwdriver into the fuse holder slot, and rotating the slot counter-clockwise. The fuse and its carrier will then pop out. FUSE RATINGS The following table lists the required fuses for standard units: Recommended Replacement Part Nominal Fuses...
FRONT PANEL CONTROLS 1. POWER Switch . This is the main power switch. When turning the instrument on, there is normally a delay of 10 seconds before anything is shown on the main display, as the internal operating system boots up. If the main menu does not appear after 30 seconds, turn off the instrument and leave it off for at least 60 seconds before applying power again.
Page 16
5. LIQUID CRYSTAL DISPLAY (LCD) . This LCD is used in conjunction with the keypad to change the instrument settings. Normally, the main menu is displayed, which lists the key adjustable parameters and their current values. The “Programming Manual for -B Instruments” describes the menus and submenus in detail.
REAR PANEL CONTROLS GATE TRIG RS-232 GPIB Note: some connectors may be in different positions than shown above, depending on the exact combination of options ordered. 1. AC POWER INPUT . An IEC-320 C14 three-pronged recessed male socket is provided on the back panel for AC power connection to the instrument. One end of the detachable power cord that is supplied with the instrument plugs into this socket.
Page 18
When triggering externally, the instrument can be set such that the output pulse width tracks the pulse width on this input, or the output pulse width can be set independently. 6. GPIB Connector . A standard GPIB cable can be attached to this connector to allow the instrument to be computer-controlled.
GENERAL INFORMATION BASIC PULSE CONTROL This instrument can be triggered by its own internal clock or by an external TTL trigger signal. In either case, two output channels respond to the trigger: OUT and SYNC. The OUT channel is the signal that is applied to the load. Its amplitude and pulse width are variable.
Page 20
> 50 ns TRIG TTL LEVELS (external input) (0V and 3V-5V) PROPAGATION DELAY (FIXED) 100 ns, FIXED SYNC OUT 3V, FIXED DELAY > 0 PULSE WIDTH AMPLITUDE, VARIABLE MAIN OUTPUT Figure C As before, if the delay is negative, the order of the SYNC and OUT pulses is reversed. The last figure illustrates the relationship between the signal when an external TTL-level trigger is used in the PW mode.
TRIGGER MODES This instrument has four trigger modes: Internal Trigger: the instrument controls the trigger frequency, and generates the clock internally. External Trigger: the instrument is triggered by an external TTL-level clock on the back-panel TRIG connector. Manual Trigger: the instrument is triggered by the front-panel “SINGLE PULSE” pushbutton.
PREVENTING OUTPUT STAGE FAILURE The output stage is protected against overload conditions by an overload circuit and fuses on the main frame back panel. However, the output switching elements may fail if the unit is triggered at a PRF exceeding 100 kHz or at duty cycles resulting in an average output power in excess of 50 Watts.
250V of voltage and 50W of power. 3. Turn on the AVR-3HF-B. The main menu will appear on the LCD. 4. To set the AVR-3HF-B to trigger from the internal clock at a PRF of 1 kHz:...
Page 24
The arrow pointer should be pointing at the frequency menu item. If it is not, press the MOVE button until it is. Press the CHANGE button. The frequency submenu will appear. Rotate the ADJUST knob until the frequency is set at 1 kHz. ...
Page 25
Press MOVE until the arrow pointer is pointing at the “ON” choice. Press CHANGE to return to the main menu. 9. To change the output amplitude: Press the MOVE button until the arrow pointer is pointing at the amplitude menu item.
PROGRAMMING YOUR PULSE GENERATOR KEY PROGRAMMING COMMANDS The “Programming Manual for -B Instruments” describes in detail how to connect the pulse generator to your computer, and the programming commands themselves. A large number of commands are available; however, normally you will only need a few of these.
ALL PROGRAMMING COMMANDS For more advanced programmers, a complete list of the available commands is given below. These commands are described in detail in the “Programming Manual for -B Instruments”. (Note: this manual also includes some commands that are not implemented in this instrument.
There are no user-adjustable internal circuits. For repairs other than fuse replacement, please contact Avtech (info@avtechpulse.com) to arrange for the instrument to be returned to the factory for repair. Service is to be performed solely by qualified service personnel.
MAINTENANCE REGULAR MAINTENANCE This instrument does not require any regular maintenance. On occasion, one or more of the four rear-panel fuses may require replacement. All fuses can be accessed from the rear panel. See the “FUSES” section for details. CLEANING If desired, the interior of the instrument may be cleaned using compressed air to dislodge any accumulated dust.
WIRING DIAGRAMS WIRING OF AC POWER (STANDARD UNITS) M ains circuits - haz ardous live . A3 - BLACK Do not attempt any repairs on this instrument POWER SWIT CH SW325-ND (CW INDUSTRIES GRS-4022-0013) beyond the fuse replacement procedures described NORMALLY: ECP180PS24 in the manual .
WIRING OF AC POWER (UNITS WITH -XP2 OPTION) M ains circuits - haz ardous live . A3 - BLACK Do not attempt any repairs on this instrument POWER SWIT CH SW325-ND (CW INDUSTRIES GRS-4022-0013) beyond the fuse replacement procedures described in the manual .
WIRING OF DC POWER M OUNT WIT H 1/2" ST ANDO FFS P1, N1 = 15000uF, 35V x 2 R1,R2 (O N BOT T OM ): 1.8K O Y CHS GND G RN T O AC/DC CONVERT ER PCB 179B, TRIM MED EXT RA LINES O N HIG H-PO WER ALL UNIT S...
PCB 156D - POLARITY CONTROL BOARD (-PN UNITS ONLY) R2 8 R2 3 "UV- ON" IS HIGH IF: 1) "UV+ HIGH" IS LOW, AND +1 5 V VC C VC C 2) "POL" IS LOW. R1 9 R1 2 0 . 1 u F R1 6 5 6 K SU RF AC E MOUN T , ON B OTTOM.
PCB 104F - KEYPAD / DISPLAY BOARD, 3/3 VC C VC C 0 . 1 u F 2 . 2 u F GN D VC C VC C VC C SD A SD A SC L SC L GN D PA D3 PA D4 PC F8 5 7 4 AN (MU ST HA VE "A"...
MAIN WIRING (STANDARD -P UNITS) G ND (NO OLO) USE 1/2" ST ANDOFFS +15V NSY VPRF OP1B M AIN BOARD, PCB 295E VSPARE N4-N6: UNUSED P4-P6: 33uF,400V (P7519-ND) R9-R10: UNUSED SYNC R7-R8: 100K O Y CONN2 SYNC O LIM EX PCB 183A-P +15V ON/OFF PG-P...
MAIN WIRING (-N UNITS WITH -XP2 OPTION) GND (NO OLO) USE 1/2" ST ANDOFFS +1 5V NSY VPRF P4-P6: UNUSED OP1B MAIN BOARD, PCB 295 E VSPARE N4-N6: 33uF, 400V (P7519-ND) +5 V R7,R8 (BOT ): UNUSED SYNC R9-R10 (BOT ): 100K OY CONN1 PCB 1 83A-N SYNC...