Avtech product is found to be defective, Avtech shall at its option repair or replace said defective item. This warranty does not apply to units which have been dissembled, modified or subjected to conditions exceeding the applicable specifications or ratings.
TABLE OF CONTENTS WARRANTY........................2 TECHNICAL SUPPORT....................2 TABLE OF CONTENTS....................3 INTRODUCTION.......................5 INCLUDED OPTIONS......................5 SPECIFICATIONS......................6 REGULATORY NOTES....................7 FCC PART 18.......................... 7 EC DECLARATION OF CONFORMITY..................7 DIRECTIVE 2011/65/EU (RoHS).....................8 DIRECTIVE 2002/96/EC (WEEE)....................8 FIRMWARE LICENSING......................9 INSTALLATION......................10 VISUAL CHECK........................10 POWER RATINGS........................ 10 CONNECTION TO THE POWER SUPPLY................10 PROTECTION FROM ELECTRIC SHOCK................11 ENVIRONMENTAL CONDITIONS..................12 LABVIEW DRIVERS......................
The voltage waveform is pulsed, similar to the main output. It is not a DC output. The AVR-G1-B-P-EMRA-R5-RP-IM includes an internal trigger source, but it can also be triggered or gated by an external source. A front-panel pushbutton can also be used to trigger the instrument.
SPECIFICATIONS Model: AVR-G1-B-P-EMRA-R5-RP-IM Amplitude: 0 to +240 Volts Pulse width (FWHM) 3 us to 30 us Load impedance: ≥ 1.6 kΩ, ≤ 900 pF. 150 mA maximum output current. Output impedance: 50 Ω (i.e., internal resistance in series with the output).
If interference is observed, check that appropriate well-shielded cabling is used on the output connectors. Contact Avtech (info@avtechpulse.com) for advice if you are unsure of the most appropriate cabling. Also, check that your load is adequately shielded. It may be necessary to enclose the load in a metal enclosure.
< 1000 ppm (0.1% by mass) DIRECTIVE 2002/96/EC (WEEE) European customers who have purchased this equipment directly from Avtech will have completed a “WEEE Responsibility Agreement” form, accepting responsibility for WEEE compliance (as mandated in Directive 2002/96/EC of the European Union and local...
Article 9 of Directive 2002/96/EC. Customers who have purchased Avtech equipment through local representatives should consult with the representative to determine who has responsibility for WEEE compliance. Normally, such responsibilities with lie with the representative, unless other arrangements (under Article 9) have been made.
INSTALLATION VISUAL CHECK After unpacking the instrument, examine to ensure that it has not been damaged in shipment. Visually inspect all connectors, knobs, liquid crystal displays (LCDs), and the handles. Confirm that a power cord, a GPIB cable, and two instrumentation manuals (this manual and the “Programming Manual for -B Instruments”) are with the instrument.
5. Do not attempt any repairs on the instrument, beyond the fuse replacement procedures described in this manual. Contact Avtech technical support (see page 2 for contact information) if the instrument requires servicing. Service is to...
6. no pollution or only dry, non-conductive pollution. LABVIEW DRIVERS A LabVIEW driver for this instrument is available for download on the Avtech web site, at http://www.avtechpulse.com/labview. A copy is also available in National Instruments' Instrument Driver Library at http://www.natinst.com/.
FUSES This instrument contains four fuses. All are accessible from the rear-panel. Two protect the AC prime power input, and two protect the internal DC power supplies. The locations of the fuses on the rear panel are shown in the figure below: Fuses #1 and #2 Fuse #4 Fuse #3...
DC FUSE REPLACEMENT The DC fuses may be replaced by inserting the tip of a flat-head screwdriver into the fuse holder slot, and rotating the slot counter-clockwise. The fuse and its carrier will then pop out. FUSE RATINGS The following table lists the required fuses: Recommended Replacement Part Nominal Fuses...
FRONT PANEL CONTROLS 1. POWER Switch . This is the main power switch. When turning the instrument on, there is normally a delay of 10 seconds before anything is shown on the main display, as the internal operating system boots up. If the main menu does not appear after 30 seconds, turn off the instrument and leave it off for at least 60 seconds before applying power again.
Page 16
5. KEYPAD . Control Name Function MOVE This moves the arrow pointer on the display. CHANGE This is used to enter the submenu, or to select the operating mode, pointed to by the arrow pointer. ×10 If one of the adjustable numeric parameters is displayed, this increases the setting by a factor of ten.
REAR PANEL CONTROLS GATE TRIG RS-232 GPIB Note: some connectors may be in different positions than shown above, depending on the exact combination of options ordered. 1. AC POWER INPUT . An IEC-320 C14 three-pronged recessed male socket is provided on the back panel for AC power connection to the instrument. One end of the detachable power cord that is supplied with the instrument plugs into this socket.
Page 18
When triggering externally, the instrument can be set such that the output pulse width tracks the pulse width on this input, or the output pulse width can be set independently. 6. GPIB Connector . A standard GPIB cable can be attached to this connector to allow the instrument to be computer-controlled.
GENERAL INFORMATION BASIC PULSE CONTROL This instrument can be triggered by its own internal clock or by an external TTL trigger signal. In either case, two output channels respond to the trigger: OUT and SYNC. The OUT channel is the signal that is applied to the load. Its amplitude and pulse width are variable.
Page 20
> 50 ns TRIG TTL LEVELS (external input) (0V and 3V-5V) PROPAGATION DELAY (FIXED) 100 ns, FIXED SYNC OUT 3V, FIXED DELAY > 0 PULSE WIDTH AMPLITUDE, VARIABLE MAIN OUTPUT Figure C As before, if the delay is negative, the order of the SYNC and OUT pulses is reversed. The last figure illustrates the relationship between the signal when an external TTL-level trigger is used in the PW mode.
TRIGGER MODES This instrument has four trigger modes: Internal Trigger: the instrument controls the trigger frequency, and generates the clock internally. External Trigger: the instrument is triggered by an external TTL-level clock on the back-panel TRIG connector. Manual Trigger: the instrument is triggered by the front-panel “SINGLE PULSE” pushbutton.
OUTPUT IMPEDANCE The AVR-G1-B-P-EMRA-R5-RP-IM has an output impedance (i.e., an internal resistance in series with the output) of 50 Ohms, to provide transmission line back- matching and short-circuit protection. If the load is capacitive, the RC time constant of the output impedance and the capacitive load will tend to degrade the rise and fall times.
1 us/div. Be sure that your oscilloscope and probe setup can handle the maximum amplitude of 240V. The load resistor should be rated for at least 240V of voltage and 20W of power. 3. Turn on the AVR-G1-B-P-EMRA-R5-RP-IM. The main menu will appear on the LCD.
Page 24
4. To set the AVR-G1-B-P-EMRA-R5-RP-IM to trigger from the internal clock at a PRF of 1 kHz: The arrow pointer should be pointing at the frequency menu item. If it is not, press the MOVE button until it is.
Page 25
Press the CHANGE button. The output submenu will appear. Press MOVE until the arrow pointer is pointing at the “ON” choice. Press CHANGE to return to the main menu. 9. To change the output amplitude: Press the MOVE button until the arrow pointer is pointing at the amplitude menu item.
PROGRAMMING YOUR PULSE GENERATOR KEY PROGRAMMING COMMANDS The “Programming Manual for -B Instruments” describes in detail how to connect the pulse generator to your computer, and the programming commands themselves. A large number of commands are available; however, normally you will only need a few of these.
These commands will satisfy 90% of your programming needs. ALL PROGRAMMING COMMANDS For more advanced programmers, a complete list of the available commands is given below. These commands are described in detail in the “Programming Manual for -B Instruments”. (Note: this manual also includes some commands that are not implemented in this instrument.
There are no user-adjustable internal circuits. For repairs other than fuse replacement, please contact Avtech (info@avtechpulse.com) to arrange for the instrument to be returned to the factory for repair. Service is to be performed solely by qualified service personnel.
MAINTENANCE REGULAR MAINTENANCE This instrument does not require any regular maintenance. On occasion, one or more of the four rear-panel fuses may require replacement. All fuses can be accessed from the rear panel. See the “FUSES” section for details. CLEANING If desired, the interior of the instrument may be cleaned using compressed air to dislodge any accumulated dust.
M a ins circuits - ha z ard ou s live . Do not attem pt any repairs on this instrument bey ond the fuse replacement procedures described in the m anual. Contact Avtech if the instrument A3 - BLACK requires serv icing. Serv ice is to be performed WARNING solely by qualified service personnel.
PCB 116D - CURRENT MONITOR SMA1 SMA-VE RT 6 4 0 4 5 6 -4 CO N9 R1 6 CH OOSE T HIS R ESI S T OR T O GET +/-3V MA X OU T U1 3 B U1 3 A ZER O (N OT U SED O N -EM R A UN ITS ) R1 7 8 2 0...
PCB 104F - KEYPAD / DISPLAY BOARD, 3/3 VC C VC C 0 . 1 u F 2 . 2 u F GN D VC C VC C VC C SD A SD A SC L SC L GN D PA D3 PA D4 PC F8 5 7 4 AN (MU ST HA VE "...