Using An External Pump - Fluke 718-100US User Manual

Pressure calibrator
Table of Contents

Advertisement

718 Series
Users Manual

Using an External Pump

XW Warning
To avoid damage to the Calibrator and possible
release of pressure, do not connect the internal
sensor to an external pressure source that
exceeds the maximum rated pressure.
To develop higher pressure or vacuum, use an external
pump (such as the Fluke Model 700PTP). Use a Fluke
Pressure Module connected to the pressure module input
on the Calibrator. Pressure modules are listed in Table .
Make overall connections as shown in Figure 6.
Refer to setup and operating instructions included with the
pressure module and pump.