5.1 GENERAL
The MD spinal board is made of materials that are highly resistant to wear and to external
agents. Despite this, the use conditions require maintenance operation and even repair in
particular conditions.
Maintenance and repair operation must be recorded in a special register an
example of which is indicated in chapter 10.

5.2 MAINTENANCE

The maintenance operations that must be carried out by the user are, in order:
a) cleaning: after every use, wash with warm (max. 40 C) potable water, possibly adding a
neutral detergent (e.g. Marseille soap). Rinse and leave to dry in a shaded place, away
from direct sources of heat,
b) disinfection, when deemed necessary or after prolonged inactivity (more than 21 days),
preceded by cleaning: wipe with a clean cloth , soaked in a water solution of sodium
hypochlorite (bleach) with a 0,1% concentration (1000 ppm). When blood or other bodily
fluids are present, the recommended concentration of sodium hypochlorite is 0,5%(5000
ppm)

5.3 REPAIR

Repair shall be strictly carried out by the manufacturer.
The user is allowed to replace the parts mentioned in paragraph 3.4.2 only,
with new and original parts.