New GentleYAG Bubble Circuit Calibration Procedure Candela Corporation New GentleYAG Bubble Circuit Calibration Procedure 8503-01-0840 Candela Corporation Proprietary Page 1 of 4 Revision 01...
Page 2
New GentleYAG Bubble Circuit Calibration Procedure Candela Corporation BUBBLE CIRCUIT CALIBRATION 1. Preparing the cryogen line for testing: 1.1. Connect any GentleYAG HP delivery system to the GentleYAG laser. 1.2. Install cryogen canister into the laser per the procedure in the Operator’s Manual.
Page 3
New GentleYAG Bubble Circuit Calibration Procedure Candela Corporation 3. Canister Bubble Detector calibration procedure: 3.1. Connect a DVM to the Canister Bubble Sense Test Point on the CPU I/O PCB at TP26 (+) and TP29 (-). Verify the voltage is 7.75 VDC ± 0.5 VDC. If not, adjust R1 of the Canister Bubble PCB to get 7.75VDC on the DVM.
Page 4
New GentleYAG Bubble Circuit Calibration Procedure Candela Corporation 8503-01-0840 Candela Corporation Proprietary Page 4 of 4 Revision 01...