New GentleYAG Bubble Circuit Calibration Procedure
Candela Corporation
New GentleYAG Bubble Circuit
Calibration Procedure
8503-01-0840
Candela Corporation Proprietary
Page 1 of 4
Revision 01

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the New GentleYAG and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Candela New GentleYAG

  • Page 1 New GentleYAG Bubble Circuit Calibration Procedure Candela Corporation New GentleYAG Bubble Circuit Calibration Procedure 8503-01-0840 Candela Corporation Proprietary Page 1 of 4 Revision 01...
  • Page 2 New GentleYAG Bubble Circuit Calibration Procedure Candela Corporation BUBBLE CIRCUIT CALIBRATION 1. Preparing the cryogen line for testing: 1.1. Connect any GentleYAG HP delivery system to the GentleYAG laser. 1.2. Install cryogen canister into the laser per the procedure in the Operator’s Manual.
  • Page 3 New GentleYAG Bubble Circuit Calibration Procedure Candela Corporation 3. Canister Bubble Detector calibration procedure: 3.1. Connect a DVM to the Canister Bubble Sense Test Point on the CPU I/O PCB at TP26 (+) and TP29 (-). Verify the voltage is 7.75 VDC ± 0.5 VDC. If not, adjust R1 of the Canister Bubble PCB to get 7.75VDC on the DVM.
  • Page 4 New GentleYAG Bubble Circuit Calibration Procedure Candela Corporation 8503-01-0840 Candela Corporation Proprietary Page 4 of 4 Revision 01...