Replacement of the 15" LCD Color TFT Display
Brightness, contrast and back light
.
Fig. 5
NOTE
NOTE
Requirements
The basic unit and monitor cart are connected to each other. The Memoskop is
synchronized by the VIDEOMED DC camera.
At the MEMOSKOP, call up the SMPTE test pattern. (Technical setup)
At the Memoskop keyboard, select the brightness/contrast curve LUT 2 for both
displays.
Adjustments
At the MEMOSKOP, LUT2 must be selected for both monitors
Activate the OSD menu on the TFT display.
Scroll to the menu "Picture Setup" with the "UP" and "DOWN" key and select it with the
"Enter" key.
Scroll to the corresponding line Brightness, Contrast or Back light intensity with the "UP"
and "DOWN" key and select it with the "Enter" key.
Set Brightness, Contrast and Back light intensity to the middle value "50%" and confirm it
by pressing the "Enter" key.
Siemens AG
Medical Solutions
For the adjustment, set the OSD time to "Off"
(permanent visible OSD menu) by selecting the OSD menu
"Options" - "Onscreen Menu" - "OSD Time".
After the adjustments, set it back to the default value
(20 seconds).
For correct adjustments of the brightness, contrast and
back light intensity, the TFT Display needs a warmup phase of 20
minutes.
The MEMOSKOP brightness/contrast curve LUT2 must be
selected for both displays.
SPR2-130.841.03
Rev. 02
Page 13 of 20
06.05