Replacement of the 15" LCD Color TFT Display
Check of the automatically detected video mode
NOTE
The basic unit and the monitor cart are connected to each other;
the system is fully functional.
The Memoskop video frequency is set correctly.
NOTE
Scroll to the "Mode Info" menu and select it by pressing the "Enter" key.
Compare the displayed values, with the values in the table Tab. 1.
Mode
Clock
H Freq.
V Freq.
Rows
H Pol
V Pol
Type
PDEN
Entry
Tab. 1
If the values are different to table Tab.1, follow the description "Manual programming"
(page 16).
Siemens AG
Medical Solutions
In the normal case, the signal source and mode will be detected
automatically.
However, if needed, the programming can be done manually.
See "Manual programming"
For systems with 50Hz camera system: the video frequency has
to be programmed to 50Hz.
For systems with 60Hz camera system: the video frequency has
to be programmed to 60Hz.
50 Hz
614 x 514
13,839 MHz
16,301 kHz
50,002
324
None
None
SoG/I
H15/L11
1 Current
SPR2-130.841.03
Rev. 02
Remarks
60Hz
640 x 444
(floating value)
13,999
(floating value)
16,549 kHz
+/- 1Hz
59,524
+/- 4 rows
278
None
None
(Sync on Green /
SoG/I
Interlaced)
H15/L11
0 Current