12
Replacement of the 15" LCD Color TFT Display
Geometry
Requirements
The basic unit and the monitor cart are connected to each other;
the system is fully functional.
Fig. 4
The basic unit and monitor cart are connected to each other. The Memoskop is
synchronized by the VIDEOMED DC camera.
At the MEMOSKOP, call up the SMPTE test pattern. (Technical setup)
Check the geometry of the displayed test pattern. It should be displayed as shown
in Fig. 4. If adjustments are necessary, follow the description "Manual programming"
(page 16).
Disconnect the basic unit from the monitor cart (disconnect X10). The MEMOSKOP is in
stand alone mode (internal synchronization).
Check the geometry of the displayed test pattern. It should be displayed as shown
in Fig. 4. If adjustments are necessary, follow the description "Manual programming"
(page 16).
Siremobil Compact L
SPR2-130.841.03
Rev. 02
06.05
Page 12 of 20