GCX Mounting Assembly Operation/Installation Manual
Dräger Model Cato
Patient Monitor – Left Side on Pivot Arm / Power Supply on Right Side
The purpose of these instructions is to:
1. Describe attachment of GCX mounting hardware to the Cato anesthesia machine.
2. Describe attachment of the Siemens monitor to the GCX mounting hardware.
WARNING: USE OF MOUNTING HARDWARE AND MONITORING COMPONENTS OTHER THAN THOSE
DESCRIBED IN THIS DOCUMENT MAY RESULT IN SERIOUS INJURY DUE TO TIPPING OF THE
ANESTHESIA MACHINE.
THIS CATO MOUNTING CONFIGURATION REQUIRES THE 40KG COUNTERWEIGHT WITH
EXTENDED WHEEL ADAPTER UPGRADE SET FROM DRÄGER (PART NUMBER: 86 018 68)
Installer: When installation is completed, provide these instructions to the end user for future reference.
INSTALLATION
Tools Required: Philips head screw driver, Hex key (incl.).
Mounting the Channel to Cato Edition
Parts Required: 4" Channel, M6 x 16 mm FHMS screws (2).
Brass Plate with inserts from Dräger (M29110)
1. To attach the Channel to the Cato left side accessory track,
locate the brass plate in the accessory track, align the
mounting holes in the Channel with the threaded holes on
the brass plate then run the M5 flat head screws (2) partly
in. (Note: Fixed Stop is at the bottom of the Channel)
2. Set the height of the Channel in the accessory track and
tighten the M6 screws to hold in place.
Prepare M Series Arm
1) Install the Tilt Limiting Stop screw.
®
with Siemens SC9000 Patient Monitor
5/16-18 x ½" SHCS (1)
Accessory Track
M6 x 16 mm (2) FHMS
Fixed Stop
DU-DR-0016-SC9000 Rev A 6/4/02 GCX Corp. Page 1 of 4

Advertisement

loading

Summary of Contents for Siemens SC9000

  • Page 1 The purpose of these instructions is to: 1. Describe attachment of GCX mounting hardware to the Cato anesthesia machine. 2. Describe attachment of the Siemens monitor to the GCX mounting hardware. WARNING: USE OF MOUNTING HARDWARE AND MONITORING COMPONENTS OTHER THAN THOSE DESCRIBED IN THIS DOCUMENT MAY RESULT IN SERIOUS INJURY DUE TO TIPPING OF THE ANESTHESIA MACHINE.
  • Page 2 Install 4” Cato Channel on Right Side As described on Page 1 of this document. Install Power Supply Mounting Adapter to Siemens Power Supply Mount (Siemens P/N: 4720129) 1) Install according to the instructions included with the adapter. Slide Power Supply Mount into Channel & Insert Power Supply 1) Align the Slide of the mount with edge of the Channel insert into channel.
  • Page 3 5/32 (4 mm) hex key. Cable Management 1. An open cavity beneath the arm with flexible cable covers manage cables going to and from the front and rear of the arm. Flexible Cable Covers DU-DR-0016-SC9000 Rev A 6/4/02 GCX Corp. Page 3 of 4...
  • Page 4 Dräger P/N DESCRIPTION 4720129 Siemens Power Supply Holster Siemens Power Supply Flush Mount for SI-0041-06 MM15754 Siemens P/N 4720129 DR-0017-82 MM15719 4" Channel /w T-Nuts M29110 Accessory Track Brass Plate (Dräger) DU-DR-0016-SC9000 Rev A 6/4/02 GCX Corp. Page 4 of 4...