Appendix - HoMedics HL858DM Instruction Manual

Automatic upper arm blood pressure monitor
Table of Contents

Advertisement

Appendix

Guidance and manufacturer's declaration – electromagnetic emissions
The device is intended for use in the electromagnetic environments listed below, and should only
be used in such environments:
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environments listed below, and should only
be used in such environments:
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Power frequency
(50 or 60 Hz)
magnetic field
IEC 61000-4-8
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Compliance
Complies
IEC 60601
test level
± 8 kV contact discharge
± 15 kV air discharge
30 A/m
50 or 60 Hz
± 2 kV for power supply
lines
± 1 kV for input/output
lines
AC Power port
±1 KV Line to Line
0% UT; 0.5 cycle
At 0°,45°,90°,135°,180
°,225°,270°and 315°.
0 % UT; 1 cycles
70 % UT; 25/30 cycles
0 % UT; 250/300 cycle
Electromagnetic environment – guidance
RF energy is used only to maintain device's
operation. Therefore, its RF emissions are so low
Group 1
that it's not likely to cause any interference in
nearby electronic equipment.
The device is suitable for use in all
Class B
establishments, including domestic
establishments, and those directly connected to
Class A
the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
Compliance level
± 8 kV contact discharge
± 15 kV air discharge
30 A/m
50 or 60 Hz
± 2 kV for power supply
lines
± 1 kV for input/output
lines
AC Power port
±1 KV Line to Line
0% UT; 0.5 cycle
At 0°,45°,90°,135°,180°,225
°,270°and 315°.
0 % UT; 1 cycles
70 % UT; 25 cycles
0 % UT; 250 cycle
36
Electromagnetic environment –
guidance
In the case of air discharge testing, the
climatic conditions shall be within the
following ranges:
Ambient Temperature:15℃~35℃,
Relative Humidity:30%~60%.
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the device
requires continued operation during
power mains interruptions, it is
recommended that the device be
powered from an uninterruptible power
supply or a battery.
VER: A001 202203