Gilson MA-26X Operating Manual

Gilson MA-26X Operating Manual

Aqua-check moisture tester
Hide thumbs Also See for MA-26X:

Advertisement

Quick Links

OPERATING MANUAL
Gilson Aqua-Check Moisture Tester
MA-26X
Rev: 5/2022
PHONE: 800-444-1508
FAX: 800-255-5314
P.O. Box 200, Lewis Center, Ohio 43035-0200
740-548-7298
Email: customerservice@gilsonco.com
Product Web Page:
globalgilson.com
740-548-5314

Advertisement

loading

Summary of Contents for Gilson MA-26X

  • Page 1 OPERATING MANUAL Gilson Aqua-Check Moisture Tester MA-26X Rev: 5/2022 PHONE: 800-444-1508 FAX: 800-255-5314 P.O. Box 200, Lewis Center, Ohio 43035-0200 740-548-7298 Email: customerservice@gilsonco.com Product Web Page: globalgilson.com 740-548-5314...
  • Page 2: Table Of Contents

    Gilson Aqua-Check Moisture Tester MA-26X Safety Instructions ..........................3 1.0 Unpacking & Set Up ........................4 2.0 Introduction .............................4 3.0 Operating Instructions ........................5 4.0 Accuracy ............................8 5.0 Troubleshooting ..........................9 6.0 Accessories .............................9 Gilson Aqua-Check Moisture Tester, MA-26X...
  • Page 3: Safety Instructions

    SAFETY INSTRUCTIONS Please read these instructions thoroughly to familiarize yourself with the operation of the MA-26X before attempting to run it. Users must understand and follow local regulations and calcium carbide Material Safety Data Sheet (MSDS) guidelines and procedures regarding transportation, storage, handling, and disposal. The buyer is responsible for ensuring that users are properly trained, that they are aware of all of the information and instructions in this document, and that they are aware of the potential risks of operating the apparatus. The manufacturer will not be responsible for any damage to people and/...
  • Page 4: Unpacking & Set Up

    • Small, Fine-Bristle Brush • Sample Cup • Heavy-Duty plastic Case The MA-26X includes all of the components listed, except the Calcium Carbide Reagent*. Any components or accessories are available individually from Gilson. Model numbers to order are listed separately in this manual. *Calcium Carbide Reagent is available separately from Gilson as MAA-44 in a 10lb can. 2.0 INTRODUCTION The Gilson Aqua-Check Moisture Tester is designed to measure the water (moisture) content of soils, aggregates, and other non-reactive materials. The Aqua-Check apparatus consists of an aluminum...
  • Page 5: Operating Instructions

    It is also possible to obtain better resolution of drier materials (<10%) by doubling the sample size and halving the gauge reading. Gilson Aqua-Check Moisture Tester, MA-26X...
  • Page 6 (MSDS) included with each container of calcium carbide reagent. STEP 5: Adding calcium carbide reagent to chamber cap STEP 6. Holding the Aqua-Check Vessel horizontally, place the chamber cap on the end of the chamber. Position the black dot on the cap so that it will be facing away from your body when the pressure is released. Move the stirrup into place over the cap and tighten the top screw to seal the chamber. Use care during this step not to mix the reagent and sample before the chamber is sealed. STEP 6: Placing chamber cap on the end of STEP 6: Tightening the top screw to seal the chamber the chamber Gilson Aqua-Check Moisture Tester, MA-26X...
  • Page 7 STEP 10. Place the used sample/reagent mixture and residue in an open dry container and dispose of in accordance with local regulations or Section 13 of the MSDS. Use the Coarse-Bristle Brush to clean the vessel and prepare for the next test. STEP 9: Loosening top screw to vent pressurized gas Gilson Aqua-Check Moisture Tester, MA-26X...
  • Page 8: Accuracy

    Typical Calibration Curve Calcium Carbide Tester Gauge Reading, Calcium Carbide Tester Exceptionally cool or warm ambient temperatures may cause fluctuations in test results. Repeated testing or hot weather may heat up the vessel body, and show test results that are artificially high. Allow the vessel to cool down before further testing. In cold weather, it may help to run two or three tests in rapid succession to warm up the vessel. The maximum recommended particle size is 0.75in (20mm). Some test methods may have a smaller size limit. Gilson Aqua-Check Moisture Tester, MA-26X...
  • Page 9: Troubleshooting

    MAA-52 Heavy-Duty Waterproof Plastic Case MAA-46 ¹ Due to shipping restrictions, reagent is sold separately in 10lb (4.5kg) cans. Material from these larger cans can be used to replenish smaller containers for field use. MAA-48 Gilson Aqua-Check Moisture Tester, MA-26X...