SAFETY INSTRUCTIONS Please read these instructions thoroughly to familiarize yourself with the operation of the MA-26X before attempting to run it. Users must understand and follow local regulations and calcium carbide Material Safety Data Sheet (MSDS) guidelines and procedures regarding transportation, storage, handling, and disposal. The buyer is responsible for ensuring that users are properly trained, that they are aware of all of the information and instructions in this document, and that they are aware of the potential risks of operating the apparatus. The manufacturer will not be responsible for any damage to people and/...
• Small, Fine-Bristle Brush • Sample Cup • Heavy-Duty plastic Case The MA-26X includes all of the components listed, except the Calcium Carbide Reagent*. Any components or accessories are available individually from Gilson. Model numbers to order are listed separately in this manual. *Calcium Carbide Reagent is available separately from Gilson as MAA-44 in a 10lb can. 2.0 INTRODUCTION The Gilson Aqua-Check Moisture Tester is designed to measure the water (moisture) content of soils, aggregates, and other non-reactive materials. The Aqua-Check apparatus consists of an aluminum...
It is also possible to obtain better resolution of drier materials (<10%) by doubling the sample size and halving the gauge reading. Gilson Aqua-Check Moisture Tester, MA-26X...
Page 6
(MSDS) included with each container of calcium carbide reagent. STEP 5: Adding calcium carbide reagent to chamber cap STEP 6. Holding the Aqua-Check Vessel horizontally, place the chamber cap on the end of the chamber. Position the black dot on the cap so that it will be facing away from your body when the pressure is released. Move the stirrup into place over the cap and tighten the top screw to seal the chamber. Use care during this step not to mix the reagent and sample before the chamber is sealed. STEP 6: Placing chamber cap on the end of STEP 6: Tightening the top screw to seal the chamber the chamber Gilson Aqua-Check Moisture Tester, MA-26X...
Page 7
STEP 10. Place the used sample/reagent mixture and residue in an open dry container and dispose of in accordance with local regulations or Section 13 of the MSDS. Use the Coarse-Bristle Brush to clean the vessel and prepare for the next test. STEP 9: Loosening top screw to vent pressurized gas Gilson Aqua-Check Moisture Tester, MA-26X...
Typical Calibration Curve Calcium Carbide Tester Gauge Reading, Calcium Carbide Tester Exceptionally cool or warm ambient temperatures may cause fluctuations in test results. Repeated testing or hot weather may heat up the vessel body, and show test results that are artificially high. Allow the vessel to cool down before further testing. In cold weather, it may help to run two or three tests in rapid succession to warm up the vessel. The maximum recommended particle size is 0.75in (20mm). Some test methods may have a smaller size limit. Gilson Aqua-Check Moisture Tester, MA-26X...
MAA-52 Heavy-Duty Waterproof Plastic Case MAA-46 ¹ Due to shipping restrictions, reagent is sold separately in 10lb (4.5kg) cans. Material from these larger cans can be used to replenish smaller containers for field use. MAA-48 Gilson Aqua-Check Moisture Tester, MA-26X...